Series 3700 System Switch/Multimeter User's Manual
Section 9: Series 3700 Module Schematics and Connections
3700S-900-01 Rev. C / July 2008
9-5
Example 2
This example is for a partially loaded Model 3706 with Model 3723 cards (all 1-
pole mode).
Slot #
Card
Channel relays closed
Backplane relays closed
SLOT 1
3723
107
1
SLOT 2
3723
107
1
SLOT 3
EMPTY
0
0
SLOT 4
3723
107
1
SLOT 5
3723
107
1
SLOT 6
EMPTY
0
0
This produces the following power consumption:
Slot 1 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 2 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 3 Power Consumed =
0
+
0
+
0
=
0
Slot 4 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 5 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 6 Power Consumed =
0
+
0
+
0
=
0
Totals for each bank can then be calculated as follows:
Slot 1
Slot 2
Slot 3
Total
Bank #1 Power Consumed =
6150
+
6150
+ 0
=
12300
Slot 4
Slot 5
Slot 6
Total
Bank #2 Power Consumed =
6150
+
6150
+ 0
=
12300
Result: Because each bank has not exceeded maximum power, this is OK.
Example 3
This example is for a fully loaded Model 3706-S with Model 3723 cards (all 2
pole mode).
Slot #
Card
Channel relays closed
Backplane relays closed
SLOT 1
3723
60
4
SLOT 2
3723
60
4
SLOT 3
3723
60
4
SLOT 4
3723
60
4
SLOT 5
3723
60
4