![background image](http://html2.mh-extra.com/html/keithley/2520/2520_service-manual_4001830067.webp)
Model 2520 Pulsed Laser Diode Test System Service Manual
Calibration
2-37
Step 8: Program calibration dates
Use the following commands to set the calibration date and calibration due date:
:CAL:PROT:DATE
<yyyy>, <mm>, <dd>
(Calibration date)
:CAL:PROT:NDUE
<yyyy>, <mm>, <dd>
(Next calibration due date)
Note that the year, month, and date must be separated by commas.
Step 9: Save calibration constants
Calibration is now complete. You can store the calibration constants in EEROM by
sending the following command:
:CAL:PROT:SAVE
NOTE
Calibration will be temporary unless you send the SAVE command.
Step 10: Lock out calibration
To lock out further calibration, send the following command after completing the
calibration procedure:
:CAL:PROT:LOCK
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......