4-14
Troubleshooting
Model 2520 Pulsed Laser Diode Test System Service Manual
Power supply checks
Power supply problems can be checked out using
for circuit theory on the power supply.
Table 4-2
Power supply checks
Step
Item/
component*
Required condition
Remarks
1
2
3
4
5
6
7
8
9
10
11
12
Line fuse
Line power
TP1
TP2
TP14
TP15
TP16
TP20
TP21
TP22
TP23
TP32
Check continuity.
Plugged into live receptacle, power on.
+25V, ±10%
-25V, ±10%
+25V, ±5%
+12V, ±5%
+5V, ±5%
+5V, ±5%
-5V, ±5%
+12V, ±5%
-12V, ±5%
+3.3V, ±5%
Remove to check.
Check for correct power-up sequence.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
Referenced to jumper W3.
* Test points (TP) are marked on circuit boards.
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......