Model 2520 Pulsed Laser Diode Test System Service Manual
Performance Verification
1-3
WARNING
While the Model 2520 does not incorporate a laser, it is designed to
operate (power) laser diode devices. Read all safety precautions listed
at the beginning of the Model 2520 User’s Manual. The following
safety practices must be used to protect operators and other users of
this product from potential exposure to laser radiation:
•
Operators must be protected from radiation and electrical
hazards at all times.
•
The installer must comply with all applicable laws and regulations
on laser safety. This requirement includes warning signs and
operator training.
•
The interlock is required for safe operation. The test fixtures must
ensure that the interlock circuit is disabled (source outputs
inhibited) so that an operator is not exposed to any radiation. The
test fixture interlock must not be defeated.
•
The testhead key control must be used to prevent operation unless
authorized by the responsible body. This requirement must be
part of the facilities administrative controls for laser safety.
Outputs cannot be turned on with the key removed.
•
When servicing the test system, any required personnel protection
equipment (e.g. laser safety goggles) must be provided by the
customer’s responsible body.
•
The customer’s laser safety officer (LSO) must review and
approve all installations before they are put into operation. Any
safety concerns must be immediately reported to the customer’s
LSO.
•
If at any time, the indicators provided on the testhead for
INTERLOCK STATUS or LASER POWER ON should fail to
light or properly indicate status, immediately contact a Keithley
service representative for repair. Failure to do so may expose the
user to hazards without proper warnings. See “Interlock status
indicator test sequence” in Section 9 of the User’s manual for
details on testing the indicator lights.
•
Maximum isolation from earth ground is 10V. Exceeding this
value may result in a shock hazard.
•
When making connections, do not leave any exposed connections.
Ensure that all external circuits are properly insulated.
NOTE
If the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the
correct course of action.
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......