background image

Section 1: Calibrating and adjusting the Model 2110 

Model 2110 5½ Digit Multimeter Calibration Manual 

 

1-2 

2110-905-01 Rev. A / July 2012 

 

Figure 1: Model 2110 KI-TOOL software 

 

 

5.  The procedure consists of connecting the Fluke 5700A/5725A and 5520A/5522A output to the 

Model 2110 front panel input, setting up appropriate ranges, and performing calibration by 

sending specific SCPI commands. These commands are entered into the Send String field 

followed by clicking the Write button. To access the command control window, click Tool > 

Command Control. 

 

 

The figure below shows the Command Control window. Ensure that the correct model number is 

selected before sending any SCPI commands. 

 

Содержание 2110-100

Страница 1: ...N O C F O E R U S A E M R E T A E R G A Model 2110 5 Digit Multimeter Calibration Manual 2110 905 01 Rev A July 2012 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipm...

Страница 2: ...ation herein in whole or in part without the prior written approval of Keithley Instruments Inc is strictly prohibited All Keithley Instruments product names are trademarks or registered trademarks of...

Страница 3: ......

Страница 4: ...ment control and data I O connections are for connection to Category I sources unless otherwise marked or described in the user documentation Exercise extreme caution when a shock hazard is present Le...

Страница 5: ...dangers that might result in personal injury or death Always read the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation expl...

Страница 6: ...justment 100 M 2 11 Current function adjustment DC and AC 2 12 DC current range adjustment 2 12 DC high current adjustment 2 13 AC current range adjustment 2 14 AC high current adjustment 2 15 Tempera...

Страница 7: ......

Страница 8: ...proper VISA layer Keithley I O Layer 5 0 or greater included on the CD ROM that came with your Model 2110 NI VISA 3 1 or greater Agilent I O Library Suite 14 2 or greater 1 Install the USB device dri...

Страница 9: ...5520A 5522A output to the Model 2110 front panel input setting up appropriate ranges and performing calibration by sending specific SCPI commands These commands are entered into the Send String field...

Страница 10: ...nnected to the appropriate input jacks Make sure the calibrator is in OPERATE before you complete each calibration step Do not connect test equipment to the Model 2110 through a scanner or other switc...

Страница 11: ...e Fluke 5725A Amplifier or the equivalent or better and the Fluke 5520A 5522A for temperature and capacitance to calibrate the Model 2110 Calibration code To unlock and initiate calibration 1 Unlock c...

Страница 12: ...in injury or death To eliminate the thermal EMFs due to the differences between two metals use copper leads to connect your source signal to the DMM DCV range adjustment 1 Set the Model 5700A for 0 V...

Страница 13: ...mode 16 Adjust the input offset by sending the following command CAL PROT DC STEP 1 0 17 Wait until the following message is displayed on the Model 2110 Cali OK 18 Set the Model 5700A output to 1 V D...

Страница 14: ...2110 Cali OK 40 Set the Model 5700A output to 100 V DC 41 Adjust full scale measurement by sending the following command CAL PROT DC STEP 2 100 42 Wait until the following message is displayed on the...

Страница 15: ...e Model 5700A output mode to STANDBY 11 Configure the Model 2110 for the 1 V range by sending the following command CONF VOLT AC 1 12 Set the Model 5700A to 0 1 V and the output mode to OPERATE 13 Adj...

Страница 16: ...5 V and the output mode to OPERATE 37 Adjust to 10 percent of range by sending the following command CAL PROT AC STEP 1 75 38 Wait until the following message is displayed on the Model 2110 Cali OK 39...

Страница 17: ...gure the Model 2110 for 2 wire 100 resistance measurements by sending the following command CONF RES 1E2 4 Set the measurement rate to 10 PLC by sending the following command RES NPLC 10 5 Disable the...

Страница 18: ...Set the measurement rate to 10 PLC by sending the following command RES NPLC 10 24 Set the Model 5700A to 0 and the output mode to OPERATE 25 Adjust input offset by sending the following command CAL P...

Страница 19: ...del 2110 Cali OK 45 Set the Model 5700A output to 1 M 46 Adjust full scale measurement by sending the following command CAL PROT DC STEP 2 ACTUAL CALIBRATOR OUTPUT 47 Wait until the following message...

Страница 20: ...ing command CAL PROT DC STEP 1 0 8 Wait until the following message is displayed on the Model 2110 Cali OK 9 Set the Model 5700A output to 100 10 Adjust full scale measurement by sending the following...

Страница 21: ...4 wire 100 k resistance measurements by sending the following command CONF FRES 1E5 32 Set the measurement rate to 10 PLC by sending the following command FRES NPLC 10 33 Set the Model 5700A to 0 and...

Страница 22: ...by sending the following command CAL PROT DC STEP 2 ACTUAL CALIBRATOR OUTPUT 56 Wait until the following message is displayed on the Model 2110 Cali OK 57 Set the Model 5700A output mode to STANDBY Oh...

Страница 23: ...to 10 PLC by sending the following command CURR NPLC 10 5 Set the Model 5700A output mode to OPERATE 6 Adjust front input offset by sending the following command CAL PROT DC STEP 1 0 7 Wait until the...

Страница 24: ...he Model 5725 and black of the Model 2110 to black of the Model 5725 Configure the Model 5700 to output current to the Model 5725A 2 Configure the Model 2110 for the 3 A DC current range by sending th...

Страница 25: ...DC STEP 2 10 26 Wait until the following message is displayed on the Model 2110 Cali OK 27 Set the Model 5700A output mode to STANDBY Press RESET on the Model 5700A front panel to reset the calibrato...

Страница 26: ...8 Wait until the following message is displayed on the Model 2110 Cali OK 9 Set the Model 5700A output mode to STANDBY 10 Set the Model 5700A output to 3 A 1 kHz 11 Adjust full scale measurement by se...

Страница 27: ...citance function adjustment 1 Set the Model 5520A 5522A to 1 uF range and the output mode to STANDBY 2 Using shielded low thermal EMF cables connect the Fluke 5700A output HI and LO to the Model 2110...

Страница 28: ...Configure the Model 2110 for the 1 mF capacitance range by sending the following command CONF CAP 1E 3 25 Set the Model 5520A 5522A output mode to STANDBY 26 Adjust input offset by sending the follow...

Страница 29: ...PROT DATE MM DD YYYY The date string must be enclosed in double quotes string Example If calibration date is 5 Apr 2012 send the following command CAL PROT DATE 04 05 2012 Calibration due date is auto...

Страница 30: ...ange Test Point Accuracy Limits Upper Limit Lower Limit 100 mV 50 0000 mV 50 010 mV 49 990 mV 90 0000 mV 90 015 mV 89 985 mV 1 V 0 500000 V 0 50001 V 0 49992 V 0 900000 V 0 90001 V 0 89988 V 10 V 5 00...

Страница 31: ...998 V 1 000000 V 75k Hz 1 007 V 0 993 V 1 000000 V 300 kHz 1 050 V 0 950 V 10 V 10 00000 V 10 Hz 10 02 V 9 98 V 10 00000 V 30 kHz 10 03 V 9 97 V 10 00000 V 75 kHz 10 07 V 9 93 V 10 00000 V 300 kHz 10...

Страница 32: ...d reading b Max lead resistance 10 of range per lead for 100 and 1 k ranges Add 1 k per lead for all other ranges DC current accuracy verification DC CURRENT ACCURACY VERIFICATION Range Test Point Acc...

Страница 33: ...099 A 2 6901 A 0 41666667 9 0000 A 60 Hz 9 0570 A 8 9430 A 0 41666667 9 0000 A 5 kHz 9 0570 A 8 9430 A Specifications valid for 2 hour warm up at 5 1 2 digits a Slow ac filter 3 Hz bandwidth b Pure si...

Страница 34: ...d Measurement rate set to 10 PLC fixed Capacitance accuracy verification CAPACITANCE ACCURACY VERIFICATION Range Test Point Accuracy Limits Upper Limit Lower Limit 1 nF 1 00000 nF 1 02800 nF 0 97200 n...

Страница 35: ...ge without notice All Keithley trademarks and trade names are the property of Keithley Instruments Inc All other trademarks and trade names are the property of their respective companies A G R E A T E...

Отзывы: