Section 1: Calibrating and adjusting the Model 2110
Model 2110 5½ Digit Multimeter Calibration Manual
1-4
2110-905-01 Rev. A / July 2012
Recommended test equipment
Use the Fluke 5700A Calibrator and the Fluke 5725A Amplifier (or the equivalent or better) and the
Fluke 5520A/5522A for temperature and capacitance to calibrate the Model 2110.
Calibration code
To unlock and initiate calibration:
1. Unlock calibration by sending the following command:
CAL:PROT:CODE
<up to 8-character string>
The default command is:
CAL:PROT:CODE 123456
2. Initiate calibration by sending the following command:
CAL:PROT:INIT