Section 2: Adjustment
Model 2110 5½ Digit Multimeter Calibration Manual
2-10
2110-905-01 Rev. A / July 2012
21. Set the Model 5700A output mode to STANDBY.
22. Configure the Model 2110 for 4-wire, 10 kΩ resistance measurements by sending the following
command:
CONF:FRES 1E4
23. Set the measurement rate to 10 PLC by sending the following command:
FRES:NPLC 10
24. Set the Model 5700A to 0Ω and the output mode to OPERATE.
25. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
26. Wait until the following message is displayed on the Model 2110:
Cali OK
27. Set the Model 5700A output to 10kΩ.
28. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
29. Wait until the following message is displayed on the Model 2110:
Cali OK
30. Set the Model 5700A output mode to STANDBY.
31. Configure the Model 2110 for 4-wire, 100 kΩ resistance measurements by sending the following
command:
CONF:FRES 1E5
32. Set the measurement rate to 10 PLC by sending the following command:
FRES:NPLC 10
33. Set the Model 5700A to 0Ω and the output mode to OPERATE.
34. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
35. Wait until the following message is displayed on the Model 2110:
Cali OK
36. Set the Model 5700A output to 100kΩ.
37. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
38. Wait until the following message is displayed on the Model 2110:
Cali OK
39. Set the Model 5700A output mode to STANDBY.
40. Configure the Model 2110 for 4-wire, 1 MΩ resistance measurements by sending the following
command:
CONF:FRES 1E6
41. Set the measurement rate to 10 PLC by sending the following command:
FRES:NPLC 10
42. Set the Model 5700A to 0Ω and the output mode to OPERATE.
43. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
44. Wait until the following message is displayed on the Model 2110:
Cali OK