Model 2110 5½ Digit Multimeter Calibration Manual
Section 2: Adjustment
2110-905-01 Rev. A / July 2012
2-17
14. Set the Model 5520A/5522A output to 10UF.
15. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,1E-5
16. Wait until the following message is displayed on the Model 2110:
Cali OK
17. Configure the Model 2110 for the 100 uF capacitance range by sending the following command:
CONF:CAP 1E-4
18. Set the Model 5520A/5522A output mode to STANDBY.
19. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
20. Wait until the following message is displayed on the Model 2110:
Cali OK
21. Set the Model 5520A/5522A output to 100 uF.
22. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,1E-4
23. Wait until the following message is displayed on the Model 2110:
Cali OK
24. Configure the Model 2110 for the 1 mF capacitance range by sending the following command:
CONF:CAP 1E-3
25. Set the Model 5520A/5522A output mode to STANDBY.
26. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
27. Wait until the following message is displayed on the Model 2110:
Cali OK
28. Set the Model 5520A/5522A output to 1 mF.
29. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,1E-3
30. Wait until the following message is displayed on the Model 2110:
Cali OK
31. Configure the Model 2110 for the 10 mF Capacitance range by sending the following command:
CONF:CAP 1E-2
32. Set the Model 5520A/5522A output mode to STANDBY.
33. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
34. Wait until the following message is displayed on the Model 2110:
Cali OK
35. Set the Model 5520A/5522A output to 10 mF.
36. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,1E-2
37. Wait until the following message is displayed on the Model 2110:
Cali OK