HP3458A Emulation Mode
H-6
Query
DELAY?
Returns delay time in seconds.
Description
The
DELAY
command programs the time interval between the trigger event and the first sample event
in seconds. Although the valid parameter range is from 1E-7s to 6000s, the Model 2002 rounds any
value below 1ms (0.001s) to 1ms. Specifying a delay of 0 selects the minimum delay time possible.
Example
DELAY 50E-3
Program 50ms delay.
DISP
Purpose
To enable/disable the front panel display, to clear the display, and to send a message to the display.
Format
DISP <control>, "<message>"
Parameter
<control>
<control> Parameter
Numeric equivalent
Description
OFF
0
Display off or message.
ON
1
Normal operation.
MSG
2
Display message.
CLR
3
Clear display.
<message>
Up to 52-character string enclosed in double quotes.
Default
ON
Query
DISP?
Returns present value of
<control>
parameter.
Description
The
DISP
command enables/disables the multimeter's display and may be used to send a message to
the display, or to clear the display. ON selects normal operation. OFF displays a message if a valid
message parameter has been sent; otherwise, all dashes are displayed. Readings are no longer dis-
played, and the display is not updated, except to service front panel keystrokes. The MSG parameter
simply displays a message on the front panel, and the CLR parameter clears the display entirely. The
<message>
parameter is any string up to 52 characters enclosed in double quotation marks.
Example
DISP OFF
Turn off display.
DISP ON
Enable normal display operation.
DISP MSG, "Model 2002"
Display message.
EMASK
Purpose
To enable certain error conditions to set the error bit in the status register.
Format
EMASK <bit_value>