![background image](http://html1.mh-extra.com/html/keithley/2002/2002_user-manual_3972071281.webp)
IEEE-488 Reference
3-158
This command is used to set the date for the RTCLock control source. Setting an invalid date
(i.e. February 29, 1994) for the clock event results in an error and the previous date is retained.
PRINT #1, "output 16; :arm:rtcl:date <1993>, <12>, <1>; date?"
' Set date to
December 1,
1993
and
query
#1,
"enter
16"
' Get response
message from
2002
:TIME <hr>, <min>, <sec>
:ARM[:SEQuence[1]][LAYer[]]:RTCLock:TIME <hr>, <min>, <sec>
Set time for RTCLock control source
<hr> =
0 to 23
Specify hour (24 hour format)
<min> =
0 to 59
Specify minute
<sec> =
0.00 to 59.99
Specify second (round to 1/100 sec.)
:DATE?
Query the time
This command is used to set the time for the RTCLock control source. Note that the time must
be set using the 24-hour format (i.e. hour 13 is 1pm). Setting an invalid time results in an error
and the previous time is retained.
:TIME? returns the time to nearest hundreth of a second.
PRINT #1, "output 16; :arm:rtcl:time <14>, <36>, <0>; time?"
' Set time to
2:36 pm and
query
PRINT #1, "enter 16"
' Get response
message from
2002
Description
Program fragment
Parameters
Query
Description
Program fragment
Содержание 2002
Страница 123: ...Front Panel Operation 2 106...
Страница 145: ...IEEE 488 Reference 3 22...
Страница 167: ...IEEE 488 Reference 3 44...
Страница 284: ...A Specifications A 1...
Страница 362: ......
Страница 364: ...This page left blank intentionally...
Страница 365: ......