IEEE-488 Reference
3-113
The :ACQuire command is then used to store the simulated reference temperature.
PRINT #1, "output 16; :unit:temp c"
' Select °C temp scale
PRINT #1, "output 16; temp:rjun1:sim 0; sim?"
' Set reference for 0°C
PRINT #1, "enter 16"
' Get response from 2002
:TCOefficient <n>
[:SENSe[1]]:TEMPerature:RJUNctionX:REAL:TCOefficient <n> Specify temperature coefficient
<n> = -0.09999 to +0.09999
Specify temperature coefficient
DEFault
+0.01 temperature coefficient
MINimum
-0.09999 temperature coefficient
MAXimum
+0.09999 temperature coefficient
:TCOefficient?
Query temperature coefficient (TC)
:TCOefficient? DEFault
Query *RST default TC
:TCOefficient? MINimum
Query lowest allowable TC
:TCOefficient? MAXimum
Query largest allowable TC
This command is used to specify the temperature coefficient (TC) of the “real” temperature ref-
erence junction. TC is specified in °C/volt and is not affected by the :UNIT :TEMPerature com-
mand. The :ACQuire command can then be used to acquire the reference temperature.
PRINT #1, "output 16; temp:rjun1:real:tco .05; tco?"
' Set temp coefficient
PRINT #1, "enter 16"
' Get response from 2002
:OFFSet <n>
[:SENSe[1]]:TEMPerature:RJUNctionX:REAL:OFFSet <n>
Specify voltage offset
<n> = -0.09999 to +0.09999
Specify voltage offset at 0°C
DEFault
0.01V offset
MINimum
-0.09999V offset
MAXimum
+0.09999V offset
:OFFSet?
Query voltage offset
:OFFSet? DEFault
Query *RST default voltage offset
:OFFSet? MINimum
Query lowest allowable voltage offset
:OFFSet? MAXimum
Query largest allowable voltage offset
This command is used to specify the offset voltage at 0°C for the specified reference junction.
PRINT #1, "output 16; temp:rjun1:real:offs .05; offs?"
' Set voltage offset
PRINT #1, "enter 16"
' Get response from
2002
:ACQuire
[:SENSe[1]]:TEMPerature:RJUNctionX:ACQuire
Update reference temperature
This action command is used to update the stored reference temperature (Simulated or Real) for
generating a temperature measurement.
PRINT #1, "output 16; ":temp:rjun1:acq"
' Update reference temperature
Program fragment
Parameters
Query
Description
Program fragment
Parameters
Query
Description
Program fragment
Description
Program fragment
Содержание 2002
Страница 123: ...Front Panel Operation 2 106...
Страница 145: ...IEEE 488 Reference 3 22...
Страница 167: ...IEEE 488 Reference 3 44...
Страница 284: ...A Specifications A 1...
Страница 362: ......
Страница 364: ...This page left blank intentionally...
Страница 365: ......