IEEE-488 Reference
3-114
:OCOMpensated <b>
[:SENSe[1]]:RESistance:OCOMpensated <b>
Control offset compensation for
Ω
2
[:SENSe[1]]:FRESistance:OCOMpensated <b>
Control offset compensation for
Ω
4
<b> = 0 or OFF
Disable offset compensated ohms
1 or ON
Enable offset compensated ohms
:OCOMpensated?
Query state of offset compensated ohms
This command is used to enable or disable the offset compensated ohms feature. Offset com-
pensation is used to cancel the effects of offset voltages (such as thermal EMFs) when making
resistance measurements. Offset compensation is explained in detail in paragraph 2.4.3 (OFF-
SETCOMP).
PRINT #1, "output 16; :res:ocom on; ocom?"
' Enable offset compensation
PRINT #1, "enter 16"
' Get response from 2002
:DTCouple commands
The following commands are used to make differential thermocouple temperature measure-
ments using the Model 1801 Nanovolt Preamp. Refer to the instruction manual for the Model
1801 for details.
:TYPe <type>
[:SENSe[1]]:TEMPerature:DTCouple:TYPE <type>
Set thermocouple type
<type> = J
Set operation for Type J thermocouple
K
Set operation for Type K thermocouple
T
Set operation for Type T thermocouple
E
Set operation for Type E thermocouple
R
Set operation for Type R thermocouple
S
Set operation for Type S thermocouple
B
Set operation for Type B thermocouple
N
Set operation for Type N thermocouple
USER
Select user-defined slope for other type of thermocouples
:TYPe?
Query DTC type
This command is used to set thermocouple type for differential thermocouple measurements
using the Model 1801 Nanovolt Preamp. Selecting one of the seven standard thermocouples (J,
K, T, E, R, S, B, or N) requires that you specify the reference temperature for the reference ther-
mocouple using the :RTEMperature command.
When using other thermocouple types (USER selected), you must specify the thermocouple
slope (coefficient) using the :USLope command.
Refer to the Model 1801 Nanovolt Preamp Instruction manual for details on making differential
thermocouple temperature measurements.
PRINT #1, "output 16; :temp:dtc:type K; type?"
' Select type K TC
PRINT #1, "enter 16"
' Get response from 2002
3.18.21
Parameters
Query
Description
Program fragment
3.18.22
Parameters
Query
Description
Program fragment
Содержание 2002
Страница 123: ...Front Panel Operation 2 106...
Страница 145: ...IEEE 488 Reference 3 22...
Страница 167: ...IEEE 488 Reference 3 44...
Страница 284: ...A Specifications A 1...
Страница 362: ......
Страница 364: ...This page left blank intentionally...
Страница 365: ......