The extension cable is terminated with a Souriau UTS-LC in-line plug.
Figure 4 Souriau in-line plug
2.2.2. Connecting the fibre data communication cable to the controller
•
Remove the dust cap from the data communication cable.
•
Carefully match up the data communication cable with the corresponding port on the
back of the controller.
•
Tighten the locking nut. This will make a secure connection.
•
Check that the green and orange lights on the controller are present (see image).
Figure 5: Front panel of Keit controller
NOTE
REMARQUE
Protect the data communication fibre from stress and tight
bends (< 80mm bend radius not allowed!). Tight bends cause
a stress at the glass surface and increase the probability of
fracture.
Protégez la fibre de communication de données des contraintes et
des courbures serrées (rayon de courbure < 80 mm interdit!). Les
courbures serrées provoquent des contraintes à la surface du
verre et augmentent la probabilité de fracture.
2.3. Installation best practice
Having cleaned the probe and taken a background spectrum, the IRmadilloNA is ready to be
inserted into the reaction/process of interest.
Probe installation is unique for each application. Safe installation of the probe into the process
is the responsibility of the user and should be assessed by a suitably trained person.
Please refer to the interface drawings of your probe and mounting system for further guidance.
If you have not been provided with interface drawings, please contact [email protected].
To ensure success with installs, it is important to consider the following requirements: