Page 16
DOC1048R (21 December 2023 17:31)
www.keit.co.uk
Back to Table of Contents
(E) Earthing point
The spectrometer body has an M6 earth attachment point below the probe. This can be used
where local codes require it in addition to earthing through the power supply cable.
(F) Probe Purge inlet
The probe purge inlet is used to purge the probe only. The connection is a
4mm or 5/32”
Legris
push-fit fitting typically fitted with a blanking plug that will need to be removed to allow
connections. Rigid or fixed piping may be used if the fitting is compatible and mechanical
stress is avoided.
(G) Body (gas) purge outlet
Allows purge gas to escape the spectrometer body. There is a single outlet located beneath
the probe for purge gas to escape.
More detail on how to purge the instrument is given in DOC1049R.
WARNING
AVERTISSEMENT
Failure to properly earth the spectrometer may cause
serious injury or death.
Une mise à la terre incorrecte du spectromètre peut causer des
blessures graves, voire la mort.
Figure 8: ASM0627-09-N back and base of the instrument
(H) Body Purge inlet
The spectrometer purge inlet is used to purge the spectrometer body. The connection is a
Legris push-fit fitting typically fitted with a blanking plug that will need to be removed to allow
connections. Rigid or fixed piping may be used if the fitting is compatible and mechanical
stress is avoided.