Appendix 1 - Main Parameters:
Here are the legends used and what they mean:
O
2
:
Oxygen reading in percentage (%)
O2R:
Oxygen reference setting. '----' means switched off or set to 0%.
T Flue:
Temperature measured by the flue gas probe in Centigrade or
Fahrenheit. It displays ‘
- - -’
if the flue probe is disconnected.
T Nett :
Nett
temperature
calculated by deducting the
AMBIENT
or
INLET
temperature from the measured
FLUE
temperature. Displays in
either Fahrenheit (
°
F) or Centigrade (
°
C) or and will display ‘
- - -
’ if
the flue probe is not connected.
CO:
Carbon Monoxide
reading displayed in ppm (parts per million). ‘
- -
-’
is displayed if there is a fault with the CO sensor or the
instrument has not set to zero correctly, switch off instrument and
try again.
C0m:
Carbon Monoxide displayed in mg/m
3
.
COn:
Carbon Monoxide in ppm normalised to the O2R value
CO
Ħ
:
Carbon Monoxide in mg/m3 normalised to the O2R value.
COa
Carbon Monoxide calculation as if there were 0% Oxygen present.
This is only displayed in ppm
CO
2
:
Carbon Dioxide calculation determined by fuel type. This is only
displayed when a combustion test is being carried out. ‘
- - -
’ is
displayed while in fresh air.
EFF :
Combustion efficiency calculation displayed in percentage either as
Gross (G) or Nett (N) or Condensing Nett (C) - Use
MENU
to
change. The calculation is determined by fuel type and uses the
calculation in British Standard BS845. The efficiency is displayed
during a combustion test, ‘
- - - -
’ is displayed while in fresh air.
X - AIR :
Excess air calculated from the measured oxygen and type of fuel
used.
Displays reading during a combustion test. ‘
- - - -
’ is displayed
while in fresh air.
Loss :
Losses calculated from Oxygen and type of fuel. Displays reading
during a combustion test. ‘
- - - -
’ is displayed while in fresh air.