Appendix 1 - Main Parameters: 

 

Here are the legends used and what they mean: 
 

O

2

 :

  

Oxygen reading in percentage (%) 

 

O2R:

 

Oxygen reference setting.  '----' means switched off or set to 0%. 

 

T Flue:

 

Temperature measured by the flue gas probe in Centigrade or 
Fahrenheit. It displays ‘

- - -’ 

if the flue probe is disconnected. 

 

T Nett :

 Nett 

temperature 

calculated by deducting the 

AMBIENT

 or 

INLET

 

temperature from the measured 

FLUE 

temperature. Displays in 

either Fahrenheit (

°

F) or Centigrade (

°

C) or and will display ‘

- - -

’ if 

the flue probe is not connected. 

 

CO: 

Carbon Monoxide

 

reading displayed in ppm (parts per million). ‘

- - 

-’  

is displayed if there is a fault with the CO sensor or the 

instrument has not set to zero correctly, switch off instrument and 
try again. 

C0m:

 

Carbon Monoxide displayed in mg/m

3

 

COn:

 

Carbon Monoxide in ppm normalised to the O2R value 

 

CO

Ħ

:

 

Carbon Monoxide in mg/m3 normalised to the O2R value. 

 

COa 

Carbon Monoxide calculation as if there were 0% Oxygen present.  
This is only displayed in ppm

 

 
CO

:

 

Carbon Dioxide calculation determined by fuel type. This is only 
displayed when a combustion test is being carried out. ‘

- - -

’ is 

displayed while in fresh air. 

 

EFF :

 

Combustion efficiency calculation displayed in percentage either as 
Gross (G) or Nett (N) or Condensing Nett (C) -  Use 

MENU

 to 

change. The calculation is determined by fuel type and uses the 
calculation in British Standard BS845. The efficiency is displayed 
during a combustion test, ‘

- - - -

’ is displayed while in fresh air. 

 

X - AIR :

 

Excess air calculated from the measured oxygen and type of fuel 
used. 

 

Displays reading during a combustion test. ‘

- - - -

’ is displayed 

while in fresh air. 

 

Loss :

 

Losses calculated from Oxygen and type of fuel. Displays reading 
during a combustion test. ‘

- - - -

’ is displayed while in fresh air. 

 

Содержание KANE200

Страница 1: ...KANE200 Efficiency Meter KANE250 Combustion Meter Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG...

Страница 2: ...G THE METER AND ITS FOUR BUTTONS 3 4 4 USING THE ROTARY DIAL STARTING FROM MENU 5 6 5 MEASURING FLUE GASES 7 8 6 WHEN YOU FINISH USING THE METER 9 7 METER PROBLEM SOLVING 9 8 METER ANNUAL RECALIBRATIO...

Страница 3: ...s The large display shows 2 readings at a time and all data can be printed via an optional infrared printer The printed data can be live data frozen data or stored data 16 sets of tests can be stored...

Страница 4: ...e Nickel Metal Hydride NMiH No other battery types are recommended WARNING The battery charger unit must only be used when NMiH batteries are fitted Alkaline batteries are not re chargeable Attempting...

Страница 5: ...the flue probe into the bottom of the meter s water trap and connect the probe s temperature plug to the socket next to the water trap check the plug s orientation is correct otherwise incorrect temp...

Страница 6: ...meter s ambient temperature will be used as the inlet temperature When countdown is finished the display s top line shows the last selected function and the bottom line displays whatever the rotary di...

Страница 7: ...eter normally operates with the pump on Press PUMP quickly to switch the pump off and on When the pump is switched off the meter displays PUMP OFF approx every 30 seconds NOTE the pump will not switch...

Страница 8: ...sults when ENTER is pressed DEL ALL clears all the memory when YES and ENTER pressed NOTE ENTER must be pressed to exit this mode VIEW the test number appears on the top line The dial position display...

Страница 9: ...te the dial back to MENU and press ENTER before you can continue In STORE View mode the rotary dial is used to select the saved parameter to be displayed Use the UP DOWN keys to change the Test number...

Страница 10: ...ircuit it displays EFF Displays calculated efficiency when O2 values are less than 18 EFFn EFFg or EFFc as selected by the user X AIR Displays calculated excess air when O2 values are less than 18 Los...

Страница 11: ...be s depth stop cone to fix it in flue diameters from 1 4 to 4 5 inch 6 to 21 mm Its maximum operating temperature is 1112 F 600 C Make sure you do not exceed the meter s operating specifications In p...

Страница 12: ...0 series manual EXAMPLE PRINTOUT IN AMBIENT AIR The standard printouts are KANE200 16 char header 16 char header Date 26 12 02 Time 14 27 08 Fuel NATU GAS O2 20 9 CO2 0 0 NETT C AMB C FLUE C EFF G LOS...

Страница 13: ...l also include the TEST NO below the header EFF G gross EFF N nett EFF C condensing KANE250 16 char Header 16 char header Date 12 26 02 Time 14 27 08 Fuel NATU GAS O2 20 9 CO2 0 0 CO ppm 00 CO a ppm 0...

Страница 14: ...s replacing Oxygen reading CO reading Meter was stored in a cold environment and is not at normal working temperature Oxygen cell or CO sensor needs replacing Pump is switched off Batteries not holdin...

Страница 15: ...AND SERVICE Although sensor life is typically more than two years the meter should be re calibrated and serviced annually to stop any long term sensor or electronics drift or accidental damage Local...

Страница 16: ...ing 1 0 reading 0 2 0 21 0 1000ppm 0 30 0 99 9 0 250 Pre programmed Fuels KANE250 Natural gas Light Oil Propane Butane LPG KANE200 as above plus heavy oil coal anthracite Dimensions Weight Handset Pro...

Страница 17: ...ble of causing interference Check all readings are as expected A level of disturbance is acceptable If not acceptable adjust the meter s position to minimize interference or switch off if possible the...

Страница 18: ...ed in mg m3 COn Carbon Monoxide in ppm normalised to the O2R value CO Carbon Monoxide in mg m3 normalised to the O2R value COa Carbon Monoxide calculation as if there were 0 Oxygen present This is onl...

Страница 19: ...Time is recorded when each combustion test is printed Note When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be ne...

Страница 20: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: