12 KANE200
series
manual
6. WHEN YOU FINISH USING THE METER
Remove its’ probe from the flue
- THE PROBE WILL BE HOT -
and let it cool. Do
not put the probe in water which will be sucked into the meter, damaging its’ pump
and sensors.
When the meter’s readings return to ambient levels, switch it off. The meter counts
down from 30 before switch off with the pump running to self clean its sensors.
7. METER
PROBLEM
SOLVING
If any problems are not solved with these solutions, contact us or an authorized repair
center.
Fault symptom
Causes / Solutions
•
Oxygen too high
•
CO2 too low
•
Air leaking into probe, tubing, water trap, connectors
or internal to meter.
•
Oxygen cell needs replacing.
•
Oxygen reading (- - - -)
•
CO reading (- - - -)
•
Meter was stored in a cold environment and is not at
normal working temperature.
•
Oxygen cell or CO sensor needs replacing.
•
Pump is switched off
•
Batteries not holding charge
•
Meter not running on mains
adapter.
•
Batteries exhausted.
•
AC charger not giving correct output.
•
Fuse blown in charger plug.
•
Meter does not respond to flue
gas
•
Particle filter blocked.
•
Probe or tubing blocked.
•
Pump not working or damaged with contaminants.
•
Net temperature or Efficiency
calculation incorrect.
•
Ambient temperature set wrong during Automatic
Calibration.
•
Flue temperature readings erratic
•
Temperature plug reversed in socket.
•
Faulty connection or break in cable or plug.
•
T flue or T nett displays (- - - -)
•
Probe not connected.
•
X-Air, EFF, COa or CO2 display
(- - - -)
•
Oxygen reading is above 18%
•
Meter just continually beeps
•
Turn dial back to MENU and press ENTER