3
*
This unit must be used with the guidance of a Physiotherapist or
Doctor.
*
Type BF equipment, Continuous Operation
*
Do not insert lead wires into a mains power supply.
*
Do not immerse unit into water or any other substance.
*
The unit is not protected from the ingress of water droplets from a
shower of rain if used outside the carrying case.
*
Do not use the Kegel8
®
Ultra unit in the presence of a fl ammable
anaesthetic gas mixture and air or with Oxygen or Nitrous Oxide.
*
If using rechargeable 9 Volt PP3 Nickel Metal Hydride batteries, be
sure to use a CE approved battery charger. Never connect the
Kegel8
®
Ultra directly to a battery charger or to any other mains
powered equipment. We advise not to use Ni-Cad rechargeable
batteries.
*
Patient Electrodes are for
single patient use only.
*
Keep out of reach of children.
*
Do not use this stimulator on your facial area unless you are under
strict guidance from a qualifi ed Clinician.
*
Application of electrodes near the thorax may increase the risk of
cardiac fi brillation.
*
Operation in close proximity (e.g. 1m) to a shortwave or microwave
therapy equipment may produce instability in the stimulator output.
*
Simultaneous connection of a patient to a high frequency surgical
equipment may result in burns at the site of the stimulator electrodes
and possible damage to the stimulator.
*
No modifi cation of this equipment is allowed!
Warnings
Содержание ULTRA
Страница 1: ...Kegel8 Ultra Operation Manual 1...
Страница 24: ...Kegel8 Ultra Operation Manual 24...
Страница 25: ...Kegel8 Ultra Operation Manual 25...
Страница 29: ...Kegel8 Ultra Operation Manual 29 Notes...
Страница 30: ...Kegel8 Ultra Operation Manual 30...
Страница 31: ...Kegel8 Ultra Operation Manual 31...