Troubleshooting

If the programme cannot be changed -

 remove the battery cover and unlock 

the unit by pressing the unlock button - see page 6 and 11.

If mA intensity cuts out  above 8 milliamps and zero mA is fl ashing on 
LCD:

1. The problem you are experiencing is most likely due to a poor connection of 
the two stainless steel electrode plates on the shaft of the probe in your body. If 
the probe is not fi tting well or momentarily becomes disconnected, for example, 
when you shift position, you will see the “0” mA symbol fl ashing on and off  and 
the current will have been cut off . Follow the tips below and try again mA button to 
increase the intensity.
2. If the internal environment is dry, it may lead to reduced electrical conductivity. 
Use a suitable, approved water based lubricant such as KY (don’t use standard 
creams or grease as the lubricant must be electrically conductive).
3.Try to pull up the pelvic fl oor by lifting up the probe and increase mA at the same 
time. This may re-establish the connection.
4. The body position which leads to lack of conductivity: The best position to 
conduct electrical stimulation using the vaginal probe is to stand up. However, with 
the shape of vaginal probes on the market, it is not ideal to stand up as the probes 
tend to fall out. We recommend that the next best position is to sit down and lean 
back slightly.
5. If you think the probe itself is not working, wash it and hold it, using your fi rst 
fi nger and thumb (or elbow crook) to make a connection across the electrode plates. 
Connect it to the stimulator as normal. Increase mA and, if the probe is functioning 
correctly, you will feel the stimulation mildly tickling in your hand (elbow crook). 
This proves the unit works when the connection is established, try again the above 
instructions. If you don’t feel any tickling in your hand, try another lead wire to 
compare (see next point), try with a diff erent probe (malfunctioning probe?). 
6. Broken lead wire. Check if the dual conductor lead wire cable is not broken, 
as it might be bent or pulled out too much, which results in no conductivity: try 
another cable. To check if the cable is good, disconnect the probe and cross the red 
and black metal pins of a lead wire, hold them fi rmly crossed with your fi ngers. 
Increase mA on the unit. If the cable conducts the electricity, the mA will go above 
10 mA and you will feel a mild tingling sensation in your fi ngers which are holding 
the crossed pins. If you feel a mild electrical current in your fi ngers when the unit 
stimulates above 10mA, this proves the unit and lead wire are not causing the mA 
intensity cut-off . Check with the probe: above point 5.
You may need to obtain another lead wire or /and a probe, please contact your 
distributor. It is a good idea to have some spare wires and probes for one user.

Содержание ULTRA

Страница 1: ...Kegel8 Ultra Operation Manual 1...

Страница 2: ...r when you report a technical fault or claim a warranty return Manufacturer s serial number of the unit Present it together with LOT number when you report a technical fault or claim a warranty return...

Страница 3: ...arger Never connect the Kegel8 Ultra directly to a battery charger or to any other mains powered equipment We advise not to use Ni Cad rechargeable batteries Patient Electrodes are for single patient...

Страница 4: ...Instructions 9 Setting up your Customized Programme 11 Lock Button 12 Continence Treatment Programmes 13 Electrode Types Tips 19 Care Maintenance Accessories and Disposal 20 Indications for use 21 Sp...

Страница 5: ...l8 Ultra is a dual channel device combining several treatment programmes into one unit Neuromuscular Stimulation is increasingly understood by Therapists and Doctors There is a better understanding of...

Страница 6: ...lly dangerous equipment Do not place electrodes Over carotid sinus nerves Over larynx or trachea Inside mouth Over the area of the heart unless so advised by your Doctor On your facial area unless und...

Страница 7: ...me SET button Reduces the intensity mA to zero and pauses the programme if a programme is running and moves the phase one step forward Displays the menu for programmes PC1 PC3 and allows the parameter...

Страница 8: ...Time of current phase Number of phases Phase mode Work Rest Alt Cont Mod 3 Hz 250 uS PC1 5 WK Min 4SEC Programme Total programme time adds cumulatively as time added to phases Set Work Rest Ramp Delay...

Страница 9: ...e to each of the pin connectors at the end of the lead wire The red and black polarity are good either way Some probes have a long wire built in with the direct socket to the unit 5 Make sure the Vagi...

Страница 10: ...eturning to storage in the plastic bag Low Battery Indicator When the battery power is low the low battery indicator will appear on the screen shown in the diagram on page 7 When the battery indicator...

Страница 11: ...est seconds between 2 and 99 seconds 9 Press CH A RP will appear and ash press CH B to set the ramp seconds between 0 1 and 9 9 seconds 10 Press CH A AL or SY will appear and ash press CH B to select...

Страница 12: ...up correctly 2 Remove the battery cover and using a thin rod gently press on the lock button as shown in the diagram on page 6 until you hear a double beep The unit is now locked and cannot be altered...

Страница 13: ...on S 150 Ramp up time secs 1 0 Ramp down time secs 0 Work time secs Cont Rest time secs 0 Alternating Synchronous Overall time 20 min Programme P02 Urge Incontinence Phase 1 Phase 2 Phase 3 Phase 4 Ph...

Страница 14: ...p time secs 1 0 Ramp down time secs 0 Work time secs 6 Rest time secs 15 Alternating Synchronous Overall time 20 min Programme P04 Stress Incontinence 2 Phase 1 Phase 2 Phase 3 Phase 4 Phase 5 Phase t...

Страница 15: ...time secs 1 0 Ramp down time secs 0 Work time secs 5 Rest time secs 5 Alternating Synchronous Overall time 20 min Programme P06 Frequency Urge 2 Phase 1 Phase 2 Phase 3 Phase 4 Phase 5 Phase time min...

Страница 16: ...ously with no rest period It is used in some countries where they have found continuous stimulation can work e ectively Programme P08 Lack of Sensitivity Phase 1 Phase 2 Phase 3 Phase 4 Phase 5 Phase...

Страница 17: ...Ramp up time secs 0 8 0 8 0 8 0 6 0 8 Ramp down time secs 0 0 0 0 0 Work time secs 5 5 5 5 5 Rest time secs 5 5 5 6 7 Alternating Synchronous Overall time 45 min Programme P10 Building Up Endurance P...

Страница 18: ...uration S 220 Ramp up time secs 1 2 Ramp down time secs 1 2 Work time secs 6 Rest time secs 10 Alternating Synchronous Overall time 20 min This programme helps to relax the pelvic muscle It may be use...

Страница 19: ...y cleansing the skin with a swab impregnated with alcohol Clip away hairy skin using scissors don t use a razor to remove the hair The electrodes conductive material is water based If it becomes satur...

Страница 20: ...ically for any discharge from the battery Remove battery completely from unit if not in use for any extended period of time typically one week Low battery indicator of 6 9 volts shown on LCD display W...

Страница 21: ...they drop onto the oor debris will adhere to conductive gel rendering the electrodes ine ective Electrode life can be considerably reduced by The type and condition of the skin Deep seated moisturiser...

Страница 22: ...accuracy 7 Time duration of the treatment selectable 1 minute to 90 minutes 8 Low Battery Indicator If the battery goes below 6 9 volts 0 2 volts the battery symbol will ash on o once every second 9 O...

Страница 23: ...RF emissions interference to be immune from e ects of interference produced by other equipment operating in their vicinity and damage due to electrostatic discharge all when operating in a typical do...

Страница 24: ...Kegel8 Ultra Operation Manual 24...

Страница 25: ...Kegel8 Ultra Operation Manual 25...

Страница 26: ...commend that the next best position is to sit down and lean back slightly 5 If you think the probe itself is not working wash it and hold it using your rst nger and thumb or elbow crook to make a conn...

Страница 27: ...for any customer service enquiries including the warranty returns Your invoice of purchase and or the rear cover of this manual should state the name and the contact details of your distributor For a...

Страница 28: ...Kegel8 Ultra Operation Manual 28 Clinical References Please contact us for any clinical referencies of Kegel8 enquiries kegel8 com...

Страница 29: ...Kegel8 Ultra Operation Manual 29 Notes...

Страница 30: ...Kegel8 Ultra Operation Manual 30...

Страница 31: ...Kegel8 Ultra Operation Manual 31...

Страница 32: ...A Document revision info Kegel8 Ultra manual English OPH200 OM EN13 23 01 2019 Savantini Ltd Savantini House Foster Street Stoneferry Road Hull HU8 8BT England Any queries should be addressed to Helpl...

Отзывы: