TonoCare Hand Unit
Electric shock protection Class II (or internally powered)
Complies with
Electrical Safety (Medical)
IEC 60601-1:2005+AMD1:2012.
IEC 60601-1-2:2014
BS EN ISO 15004-1:2009
The Mains plug is the means of isolating the device from the
mains supply – ensure the mains plug is accessible at all times.
* If connecting the TonoCare to a computer, the computer shall
comply with the requirements of EN 60601-1:2006
** Whenever the device is connected to other equipment, the
combination shall comply with the requirements of EN 60601-1:2006
IP rating IPX0
Dimensions
220 x 136 x 206mm (H x W x D)
Weight 1.044Kg
Calibrated range
5mmHg to 50mmHg
Accuracy
+/-5mmHg (95% confidence level)
Displayed accuracy
Display accuracy to 1 decimal place
e.g. 12.3
Working distance
11mm from surface of patient’s
cornea to front surface of the window.
Display
OLED 0.95”
Illumination system
LED, white and infrared
The ME Equipment includes the hand unit, the docking station,
the forehead rest and the power supply.
Docking Station
Electric shock protection Class II
IP rating IPX0
Dimensions
153 x 155 x 183mm (H x W x D)
Weight 0.725Kg
Power Supply Unit
Switch mode, multi-plug type
(110 – 240V)+/- 10%
350-700mA
Compliant to:
EN 60601-1, EN 61000-6-2,
EN 61000-6-3
Power supply output
30 VA (12V DC 2.5A)
Frequency
50/60 Hz
TonoCare & Docking Station Environmental Conditions
5.0 Specifications and electrical ratings
Operation
Storage
Transport
Temperature
limits
+10˚C to +35˚C
-10˚C to +55˚C
-40˚C to +70˚C
Humidity
limits
30% to 90%
10% to 95%
10% to 95%
Pressure
limits
800 to 1060hPa
700 to 1060hPa
500 to 1060hPa