2.3 Table Top Installation
1. Position the product package so that the arrow points
upward.
2. Use a safety knife to cut the tape sealing the box on the
top, and remove the polystyrene layer covering the package
contents.
Use caution to avoid injury from sharp edges when
handling the safety knife, and the unsealed carton edges.
3. Lift the docking station from the package and place it in a
clean area intended for the charging of the TonoCare when
not in use. Before use remove the protective film covering the
IR window on the docking station.
The TonoCare hand unit should not be used in the
proximity of the docking station.
4. Pull open the printer door, found on the left-hand side of the
docking station, and insert the roll of thermal paper provided.
Proceed to section 3.3 for further instructions on installation /
replacement of thermal paper.
5. Remove the hand unit from the package and allow the
handle to fit the lower recess in the docking station and the
measurement window to slide onto the top of the docking
station. Before use remove the protective film covering the
TonoCare display, the Front Window and the IR Window.
6. Take the forehead stabilizer from the package and allow it to
be held by its magnets at the top of the docking station, in
the area provided for it.
7. Remove the power supply from the package, plug it to the
back of the docking station and after fitting the suitable
adapter for your geography connect it to the AC inlet.
Route power cords safely to eliminate risk of tripping to
user, or damage to device.
8. The LED indicator in the hand unit should now light up to
indicate the TonoCare battery is charging.
9. Use the USB cable provided to connect the docking station to
a computer (not provided) for data export.
The computer must be compliant to
EN 60601-1:2006 (see section 5.0)