8
1.7 Warnings and cautions
(continued)
Warnings and Cautions – failure to follow these instructions may pose a risk of injury
Operation and use
(continued)
• Only use the approved Keeler accessories from the list in section
7.0, or the instrument may malfunction.
• The device will require a minimum 12 hour charge before first use.
• The device will require several 12 hour charge cycles before the
battery functions optimally.
• To prevent condensation from forming, allow instrument to come
to room temperature before use.
• Before using the TonoCare, press the Manual fire button to dispel
any minute particles of dust or moisture which may have settled
whilst the instrument was not in use.
• Check the function of the device in accordance with the
instructions in section 3.2 prior to use on the patient.
• The patient should not be in the proximity of the docking station.
• Do not touch the electrical contacts on the docking station and
the patient simultaneously.
• The forehead rest is composed of aluminium, and it is the only
part which may touch the patient. The housing of the instrument
is made of PC-ABS. Do not touch these parts if you have a known
allergy to any of the materials.
• Do not use the forehead rest in the docking station for any other
instrument, or the performance of the product may be
compromised.
• Contact of the front window / nozzle area of the TonoCare with
the eye of the patient should be avoided. If accidental contact
occurs, clean the front window and surrounding area according to
the cleaning instructions in section 4.2.
• The printer on the docking station contains a sharp serrated blade
to help cut the paper. Exercise caution to avoid contact with this
blade whenever replacing printer paper rolls or tearing printouts
from the device.
• Always verify that the printout matches the readings on the hand
unit.
Maintenance
• To maintain device performance, and ensure its safety and
effectiveness it should be serviced in accordance with the
instructions in section 4.3.
• Only decontaminate/clean in accordance with the instructions
given in section 4.2.
• If the device is not used regularly it must be recharged for at least
12 hours monthly to ensure optimum battery life.