31
4.3 Maintenance
If your TonoCare is dropped please return it to
your local authorised Keeler Service Centre.
Keeler recommends annual service of your TonoCare to
ensure the best performance of the device.
There are no user replaceable parts inside the device, including
the battery. The battery must only be replaced by trained service
personnel following the instructions in the Service Manual.
If you notice a significant reduction of the battery’s
performance, contact Keeler or your authorized distributor for
its replacement.
If the TonoCare is dropped by accident the same service centre
or distributor can verify if the device is still in calibration.
The unit performs a self-function check when switched on and
will indicate if a fault is found. A further self-test can be
activated from the menu (refer to section 3.2.1).
Do not attempt to perform any unauthorized repairs, as
this could endanger the product and patients. Do not
allow unauthorized parts to be fitted in your product.
On request, Keeler will provide necessary circuit diagrams,
component parts lists, descriptions and calibration instructions to
assist service personnel in device repair.
The MOD RECORD label on the rear of the
device is used to indicate the status of the device
in relation to significant changes.
4.4 Fault Codes
If an Error Code between 00 to 34 is seen on the screen, restart
the device and check its function. If the device does not clear
please return it to your nearest authorised Keeler Service Centre.