KCU401-2 Keeler Cryomatic Service Manual  

 

 

 

 

8. Spare Parts Listing 

 
The following spare parts are available from the distributor. 

 

Description 

Part Number 

 

 

Controller PCB Assembly 

KCU050 

Pneumatic Module Assembly 

MCU051 

Power Supply Assembly 

MCU054 

High Pressure Hose Assembly 

MCU065 

Mains Inlet 

CST261 

Case Front Assembly 

MCU068 

Case Rear 

MCU301 

Cover Plate 

MCU302 

LCD Panel 

MCU204 

Solenoid Valve N/C 0.8mm Orifice 24V/10W 

MCU205 

Solenoid Valve N/O 0.5mm Orifice 24V/14.5W 

MCU206 

Footswitch Assembly 

MCU217 

2A Anti-surge Fuse 

MIS094 

Rubber Foot 

MCU213 

1/8” Dowty Seal 

FAS014 

Inlet Filter 

MCU219 

M4 x 10 Tamperproof Pan-head Screw 

FAS055 

M3 x 16 Pozi Countersunk Screw 

FAS044 

M3 x 6 Pozi Pan-head Screw 

FAS031 

M3 x 8 Pozi Pan-head Screw 

FAS054 

M3 Shake-proof Washer 

FAS033 

M3 Plain Washer 

FAS032 

Mains Cord 

various 

Exhaust Hose (5m) 

CST267 

Exhaust Silencer 

MIS104 

Cryomatic Console Carton 

PKG015 

Cryomatic Probe Carton 

PKG004/S 

Cryomatic Instructions For Use 

KCU400 

Cryomatic Service Manual 

KCU401 

Programming Adapter 

 

Содержание Cryomatic 2509-P-1000-JPN

Страница 1: ...Cryomatic Console Service Manual for 2509 P 1000 JPN 2509 P 1000 September 2011 onwards...

Страница 2: ...et 8 Power Supply Sub assembly 9 Pneumatic System 10 Pneumatic Module Sub assembly 10 Replacing the Inlet Filter 11 Solenoid Valves 11 Electronic System 12 Footswitch 12 Controller PCB Assembly 12 LCD...

Страница 3: ...inspection Do not use hypercarbonate or phenolic based cleaning solutions or disinfectants containing cationic surfactants e g Dettox to clean the console Ensure that the system is clean and dry prio...

Страница 4: ...at the Cryomatic system has been correctly installed following the guidelines in the Cryomatic Instructions For Use Good Working Practices Potentially dangerous voltages are present inside the equipme...

Страница 5: ...nspection of mains cord for signs of damage replace if necessary c Inspection of high pressure gas hose for signs of damage replace if necessary d Inspection of footswitch for signs of damage replace...

Страница 6: ...ic from the mains electrical supply d Disconnect footswitch e Remove 4 tamper proof fixings at the rear of the enclosure using the appropriate tooling f Withdraw the front cover and carefully disconne...

Страница 7: ...e small drawer below the rocker switch on the side of the unit A screwdriver or similar should be used to unlatch the drawer to gain access to the fuse compartment Fuses should only be replaced with t...

Страница 8: ...ing manner a Remove the crimped electrical connections from the rear of the unit make a note of the correct orientation b Remove the 2 countersunk screws which are accessed from the outside of the uni...

Страница 9: ...black connection pin 4 red pin 1 5 V yellow pin 2 24 V white pin 3 12 V The entire power supply sub assembly can be removed as follows a Remove the inlet connection from the power supply b Remove the...

Страница 10: ...a Loosen and remove the bezel from the probe connector b Push back the collar and remove the flexible exhaust connection from the module c Remove the pneumatic module support bracket d Carefully remo...

Страница 11: ...e fitting is the reverse of removal A new Dowty seal should be fitted when the inlet adapter is replaced Solenoid Valves Should the need arise solenoid valves can be replaced without removing the pneu...

Страница 12: ...devices The Controller PCB Assembly can be removed in the following manner a Unlatch the connection for the LCD panel in position CN1 Carefully withdraw the flexible cable assembly b Unplug the front...

Страница 13: ...e need arise the LCD panel can be removed in the following manner a Remove the Controller PCB Assembly see previous section b Remove 4 pan head screws which secure the LCD retaining bracket Ensure tha...

Страница 14: ...ramming port is accessed at the rear of the equipment and is located behind a small removable cover Download Utility FLIP The FLIP programming application from Atmel is available as shareware This is...

Страница 15: ...Sync button the system will respond by automatically filling in the device information h Select the Device drop down menu and select Erase i On the Erase dialogue box select Full Chip Erase and press...

Страница 16: ...cognise probe a faulty memory IC in probe coupling replace re program b possible damage to electrical contacts of console or probe dirty or oxidised probe contacts inspect contacts and clean carefully...

Страница 17: ...fice 24V 10W MCU205 Solenoid Valve N O 0 5mm Orifice 24V 14 5W MCU206 Footswitch Assembly MCU217 2A Anti surge Fuse MIS094 Rubber Foot MCU213 1 8 Dowty Seal FAS014 Inlet Filter MCU219 M4 x 10 Tamperpr...

Страница 18: ...nufacturer as set forth in this warranty are expressly conditioned on the following i No alterations or repairs of any malfunction of the system shall be made to the system except by the manufacturer...

Страница 19: ...Safety Standards Complies with EN60601 1 UL60601 1 CAN CSA C22 2 No 601 1 Equipment Classification Class 1 Type BF Applied Part Mode of operation Continuous Protection against ingress Console IPx0 Fo...

Страница 20: ...Appendix 1 Console General Arrangement...

Страница 21: ...KCU401 2 Keeler Cryomatic Service Manual Page 21 of 24...

Страница 22: ...KCU401 2 Keeler Cryomatic Service Manual Page 22 of 24...

Страница 23: ...KCU401 2 Keeler Cryomatic Service Manual Page 23 of 24 Appendix 2 Pneumatic Module General Arrangement...

Страница 24: ...Berkshire SL4 4AA England Tel 44 0 1753 857177 Fax 44 0 1753 827170 www keeler co uk Keeler Instruments Inc 456 Parkway Lawrence Park Industrial Estate Broomall PA 19008 Phone 610 353 4350 800 523 56...

Отзывы: