KCU401-2 Keeler Cryomatic Service Manual
2. Scheduled Maintenance
The Cryomatic system should ideally be inspected and
serviced annually by Keeler trained personnel.
The following service points should be covered:
a)
General inspection of console for any signs of
damage.
b)
Inspection of mains cord for signs of damage –
replace if necessary.
c)
Inspection of high pressure gas hose for signs of
damage – replace if necessary.
d)
Inspection of footswitch for signs of damage –
replace if necessary.
e)
Inspection of exhaust hose assembly for signs of
damage – replace if necessary.
f)
Replace gas inlet filter.
g)
Check probe couplings for signs of wear or damage.
h)
Check probe coupling electrical contacts – clean if
necessary.
i)
Performance check using test probe.