KCU401-2 Keeler Cryomatic Service Manual
4. Pneumatic System
Pneumatic Module Sub-assembly
The complete pneumatic module sub-assembly can be
removed as follows.
a)
Loosen and remove the bezel from the probe
connector.
b)
Push back the collar and remove the flexible exhaust
connection from the module.
c)
Remove the pneumatic module support bracket.
d)
Carefully remove the cap-head screws that secure
the module to the rear enclosure.
e)
Withdraw the pneumatic module from the enclosure.
f)
Re-fitting is the reverse of removal.