KCU401-2 Keeler Cryomatic Service Manual
7. Faultfinding Chart
Problem/Fault
Possible Cause
Corrective Action
Section
1
system does not
power up
a
mains fuse in plug blown
replace with correct rating
and type
b
inlet fuse(s)
replace with correct rating
and type
3
c
mains inlet module faulty
check for mains voltage on
output terminals – replace if
necessary
3
d
power supply faulty
Check for correct voltages
on output connector –
replace if necessary
3
2
system appears to
power up but LCD
screen is blank
a
contrast setting incorrect
adjust contrast level to
correct setting
5
b
LCD module faulty
check that LCD flat ribbon is
connected properly –
replace if necessary
5
3
system appears to
power up but
nothing happens
a
firmware corrupted
re-install firmware
6
b
faulty controller PCB
replace controller PCB
5
4
no gas supply
indicated on LCD
screen
a
gas cylinder not connected
or valve not open
connect cylinder and open
valves
b
insufficient cylinder pressure replace cylinder
c
inlet filter clogged
replace inlet filter
4
d
faulty pressure sensor
replace pneumatic module
4
5
system will not
recognise probe
a
faulty memory IC in probe
coupling
replace/re-program
b
possible damage to
electrical contacts of
console or probe - dirty or
oxidised probe contacts
inspect contacts and clean
carefully if necessary
6
probe does not
freeze properly
(low or zero flow)
a
residual moisture in probe
following sterilisation
re-purge probe
b
probe blocked with
particulate matter
replace/repair probe
c
gas supply insufficient
check cylinder, connection
and valve – replace cylinder
if necessary
d
faulty/blocked control valve
replace pneumatic module
4
e
exhaust hose is blocked or
occluded.
check the exhaust hose for
blockages or occlusions -
replace if necessary
f
faulty pressure sensor
replace pneumatic module
4
g
faulty controller PCB
replace controller PCB
5
7
probe defrosts
slowly
a
probe fault – possible
internal leak
replace probe
b
faulty/leaking control valve
replace valve or pneumatic
module
4
c
leak internal to console
check for internal gas leaks
– repair leak(s) or replace
pneumatic module
4
8
excessive flow rate
from exhaust
a
faulty/leaking control valve
replace valve or pneumatic
module
4
9
gas leaks from
coupling when
probe is
disconnected
a
faulty ‘o’ rings in QR
connectors
replace pneumatic module
4
10
mute control does
not function
a
faulty membrane label
replace enclosure front
2
Содержание Cryomatic 2509-P-1000-JPN
Страница 1: ...Cryomatic Console Service Manual for 2509 P 1000 JPN 2509 P 1000 September 2011 onwards...
Страница 20: ...Appendix 1 Console General Arrangement...
Страница 21: ...KCU401 2 Keeler Cryomatic Service Manual Page 21 of 24...
Страница 22: ...KCU401 2 Keeler Cryomatic Service Manual Page 22 of 24...
Страница 23: ...KCU401 2 Keeler Cryomatic Service Manual Page 23 of 24 Appendix 2 Pneumatic Module General Arrangement...