-10-
Programming
Zircom allows users to program up to 100 courses with maximum of 20 STEPS per program.
For each STEP, set temperature, time, table position, and buzzer type.
※
Courses with (
) are not adjustable.
Programmable parameters and program instructions
Parameter
Programmable Values
Default
Course Name
Character Input
:
Alphabet & Number
Character Max
:
16 characters
Temperature
0
~
1600
℃
(2912 °F)
0°C
Time
0
:
00
~
99
:
59
(
hour
:
min.
)
0
:
00
Table Position
0
~
5
0 : Highest position
5 : Lowest position (approx 5cm from the top)
※
When muffle temperature is 900
℃
or above, the table moves to [0]
automatically.
0
Buzzer Type
0
~
5
5 types of sound.
※
0 is no sound.
0
※
Table Position
Table position affects temperature inside the muffle. Position 0 completely seals the furnace to quickly rise furnace
temperature and position 2, 3, 4, and 5 give space between the furnace and the table to quickly decrease the furnace
temperature. We recommend you to set the table at the appropriate position. For cooling stages, we strongly
recommend you to program the table position.
Program Mode (F1 Key)
Program mode allows users to input firing program.
During Standby mode, press the [F1] key to enter the Program Mode.
Key Function
F1 (Back)
:
Go back to Standby Mode.
F2 (
)
:
Change Setting Page. (
「
Temperature
・
Time
」⇔「
Table Position
・
Chime
」
)