![KCI PREVENA PLUS Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/kci/prevena-plus/prevena-plus_instructions-for-use-manual_3780894033.webp)
33
Table 202
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The PREVENA PLUS™ 125 Therapy Unit is intended for use in the electromagnetic environment specified below.
The customer or the end user of the unit should assure it is used only in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Intended Electromagnetic
Environment
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
100 kHz repetition
frequency
±1 kV for input/output
lines
100 kHz repetition
frequency
±2 kV for power supply
lines
100 kHz repetition
frequency
±1 kV for input/output
lines
100 kHz repetition
frequency
Mains power quality should be that
of a typical commercial or home
healthcare environment.
Surge
IEC 61000-4-5
±1 kV differential mode
(line - line)
±2 kV common mode
(line - earth)
±1 kV differential mode
(line - line)
±2 kV common mode
(line - earth)
Mains power quality should be that
of a typical commercial or home
healthcare environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Dips: 0% Ut for 1 cycle
70% Ut for
25 cycles at 50 Hz or
30 cycles at 60 Hz
Single phase: at 0˚
Interruptions: 0% Ut for
250 cycles at 50 Hz or
300 cycles at 60 Hz
Dips: 0% Ut for 1 cycle
70% Ut for
25 cycles at 50 Hz or
30 cycles at 60 Hz
Single phase: at 0˚
Interruptions: 0% Ut for
250 cycles at 50 Hz or
300 cycles at 60 Hz
Product has internal battery backup.
If the user of the PREVENA
PLUS™ 125 Therapy Unit requires
continued operation during
power mains interruptions, it is
recommended that the unit be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50Hz/60Hz) magnetic
field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or home healthcare
environment.
NOTE: Ut is the A.C. mains voltage prior to application of the test level.
Содержание PREVENA PLUS
Страница 381: ...EL PREVENA PLUS PREVENA PLUS CUSTOMIZABLE SENSAT R A C...
Страница 382: ...KCI KCI KCI KCI KCI KCI KCI...
Страница 384: ...408 408 408 PREVENA PLUS 125 409 410 PREVENA PLUS 410 PREVENA PLUS 125 411 416 417 495...
Страница 391: ...391 1 2 3 4 1 5 PREVENA PLUS CUSTOMIZABLE 8 cm 3 1...
Страница 392: ...392 6 7 8 9 2 10...
Страница 393: ...393 1 2 3 4 5 PREVENA PLUS CUSTOMIZABLE 8 cm 3 1...
Страница 394: ...394 6 7 8 1 2 3 4 5 PREVENA PLUS CUSTOMIZABLE 8 cm 3...
Страница 395: ...395 1 6 7 8 PREVENA PLUS CUSTOMIZABLE 3 cm 1 2 1 KCI KCI 3 cm 1 2 2 1 KCI 1 KCI 3 KCI KCI 3 cm 1 2 4 1 KCI 5 2 KCI...
Страница 396: ...396 6 7 KCI KCI 1 2 3...
Страница 399: ...399 1 PREVENA PLUS PREVENA PLUS CUSTOMIZABLE 2 PREVENA PLUS 125 LED 3 4 SENSAT R A C 3 X...
Страница 400: ...400 5 6 7 PREVENA PLUS 125 KCI LED 24 LED LED 4 2 1 8...
Страница 401: ...401 PREVENA PLUS PREVENA PLUS 125 15...
Страница 402: ...402 ID LED 15 LED 15 LED LED LED 15...
Страница 403: ...403 1 2 3 KCI 1 2 3 2 2 2 2...
Страница 404: ...404 PREVENA PREVENA PLUS 125 3 X...
Страница 408: ...408 PREVENA PLUS 125 PREVENA PLUS 1 2 3 1 2 3 PREVENA PLUS CUSTOMIZABLE...
Страница 409: ...409 PREVENA PLUS 125 KCI KCI KCI KCI KCI PREVENA PLUS 25 110 kPa 27 5 kPa 110 kPa PREVENA PLUS PREVENA PLUS 125 KCI KCI...