![KCI PREVENA PLUS Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/kci/prevena-plus/prevena-plus_instructions-for-use-manual_3780894013.webp)
13
DRESSING APPLICATION FOR NON-LINEAR INCISIONS
CAUTION: If the dressing covers the umbilicus, the umbilicus must first be fully filled with an
anti-microbial petroleum gauze prior to dressing application.
1.
Open the sterile dressing package, remove and unfold foam dressing using aseptic technique. Do
not use if package has been torn or the sterile seal has been compromised.
2.
Use enclosed ruler to measure incision before cutting.
3.
Cut foam dressing to the length and shape necessary to cover the incision.
CAUTION: Do not cut the foam over the wound, as fragments may fall into the wound. Away
from wound site, rub foam edges to remove any fragments or loose particles that may fall
into or be left in the wound upon dressing removal.
4.
Make straight cuts at the desired angles to
ensure the foam dressing pieces and seals make
a flush/even connection with no gaps.
CAUTION: Gaps at foam junctions may
potentially cause erythema due to pinching of
the skin when negative pressure is applied.
5.
Use a sealing strip to bridge where the foam
dressing pieces meet. Sealing strip should have
equal overlap on both foam dressing pieces.
NOTE: Use the PREVENA PLUS™ CUSTOMIZABLE™
Dressing to size the sealing strip as needed to
fit the width of the dressing, approximately 3
in (8 cm).
•
Liner 1 must be removed from the dressing
before applying sealing strips.
•
Place sealing strip on end of foam dressing
(creating a continuous seal). The strip should be
placed half on dressing and half off dressing and
overlap the adhesive seal on each side of the
dressing. This ensures a seal all around the edge
of the dressing.
NOTE: Having exposed adhesive material around all edges of the dressing provides a greater
seal when the drape is applied.
•
Remove the remaining liner from the sealing strip. Repeat strip application on other end of
foam dressing.
Содержание PREVENA PLUS
Страница 381: ...EL PREVENA PLUS PREVENA PLUS CUSTOMIZABLE SENSAT R A C...
Страница 382: ...KCI KCI KCI KCI KCI KCI KCI...
Страница 384: ...408 408 408 PREVENA PLUS 125 409 410 PREVENA PLUS 410 PREVENA PLUS 125 411 416 417 495...
Страница 391: ...391 1 2 3 4 1 5 PREVENA PLUS CUSTOMIZABLE 8 cm 3 1...
Страница 392: ...392 6 7 8 9 2 10...
Страница 393: ...393 1 2 3 4 5 PREVENA PLUS CUSTOMIZABLE 8 cm 3 1...
Страница 394: ...394 6 7 8 1 2 3 4 5 PREVENA PLUS CUSTOMIZABLE 8 cm 3...
Страница 395: ...395 1 6 7 8 PREVENA PLUS CUSTOMIZABLE 3 cm 1 2 1 KCI KCI 3 cm 1 2 2 1 KCI 1 KCI 3 KCI KCI 3 cm 1 2 4 1 KCI 5 2 KCI...
Страница 396: ...396 6 7 KCI KCI 1 2 3...
Страница 399: ...399 1 PREVENA PLUS PREVENA PLUS CUSTOMIZABLE 2 PREVENA PLUS 125 LED 3 4 SENSAT R A C 3 X...
Страница 400: ...400 5 6 7 PREVENA PLUS 125 KCI LED 24 LED LED 4 2 1 8...
Страница 401: ...401 PREVENA PLUS PREVENA PLUS 125 15...
Страница 402: ...402 ID LED 15 LED 15 LED LED LED 15...
Страница 403: ...403 1 2 3 KCI 1 2 3 2 2 2 2...
Страница 404: ...404 PREVENA PREVENA PLUS 125 3 X...
Страница 408: ...408 PREVENA PLUS 125 PREVENA PLUS 1 2 3 1 2 3 PREVENA PLUS CUSTOMIZABLE...
Страница 409: ...409 PREVENA PLUS 125 KCI KCI KCI KCI KCI PREVENA PLUS 25 110 kPa 27 5 kPa 110 kPa PREVENA PLUS PREVENA PLUS 125 KCI KCI...