7 Processing steps in accordance with ISO 17664 | 7.7 Packaging
24 / 30
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close
the front door and start the service procedure.
Option 2: After three minutes with no service procedure running, the device auto-
matically switches back to normal service mode.
See also:
2
Servicing with KaVo QUATTROcare PLUS
7.7 Packaging
Note
The sterilization bag must be large enough for the handpiece to fit without stretching
the bag.
The quality and use of the sterilization packaging must comply with applicable stan-
dards and be suitable for the sterilization procedure!
▶ Seal each medical device individually in a sterilization item package.
7.8 Sterilization
Sterilization in a steam sterilizer (autoclave) in accordance with ISO
17665-1
NOTICE
Improper service and care.
Premature wear and reduced product service life.
▶ Before each sterilization cycle, service the medical device with KaVo care prod-
ucts.
NOTICE
Contact corrosion due to moisture.
Damage to the product.
▶ Remove the product from the steam sterilizer immediately after the sterilization
cycle.
The KaVo medical device has a maximum temperature resistance of up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the following
sterilization processes:
▪ Sterilizer with pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
– Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
– Drying time: 30 min.
▶ Use in accordance with the manufacturer's Instructions for Use.