7 Processing steps in accordance with ISO 17664 | 7.3 Cleaning
▶ Check the patency of the spray tube and clean it using the nozzle needle (Mat. no.
0.410.0931).
▶ Rinse the spray tube and hose with at least 20 ml demineralized water using a
disposable syringe.
▶ If the spray tube is not patent after the manual rinsing procedure, the medical de-
vice must be replaced.
Brush the spray tube under running tap water for at least 20 seconds using a
medium-hard toothbrush.
The non-fixing preliminary cleaning is a central constituent and must be carried out
prior to the automated processing.
In the cleaning and disinfecting device, validated internal cleaning of the spray tube
necessitates preliminary non-fixing cleaning.
7.3 Cleaning
Note
Applicable only to the S10 S (using spray clip and hose) during automated external
and internal cleaning and disinfection:
Prior to automated processing, the spray clip must be connected to the handpiece
using the hose.
▶ Do not plug the spray clip onto the handpiece.
NOTICE
Never process this medical device in an ultrasonic device.
Functional damage and property damage.
▶ Clean manually or in a washer disinfector only.
7.3.1 Manual external cleaning
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
19 / 30