4 Startup and shut-down | 4.1 Checking the amount of water for the S10 S
12 / 30
4 Startup and shut-down
WARNING
Hazard from nonsterile products
Infection danger to the care provider and patient.
▶ Before first use and after each use, sterilize the medical device.
WARNING
Dispose of the product in the appropriate manner.
Infection hazard.
▶ Before disposal, process and sterilize the product and accessories appropriately.
NOTICE
Damage from contaminated and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
▶ Make sure that the supplied cooling air is dry, clean and free of contamination in
accordance with ISO 7494-2.
4.1 Checking the amount of water for the S10 S
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Thermal damage to the dental pulp.
▶ Adjust the water amount for the spray cooling to a minimum of 50 ml/min!
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage the
tooth.
▶ Check the spray water channels and clean the spray tube with the nozzle needle
(Mat. no. 0.410.0931) according to need.