Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF
1.011.6732
7 Processing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
18 / 30
7 Processing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from incorrectly processed products.
Contaminated products are associated with an infection risk.
▶ Take suitable personal protective measures.
▶ Remove the tool from the medical device.
▶ Take the spray clip off.
▶ Remove all residual cement, composite or blood immediately.
▶ Process the medical device as soon as possible after treatment.
▶ The medical device must be dry when transported to processing.
▶ Do not immerse in solutions or the like.
7.2 Non-fixing preliminary cleaning of the spray tube
Accessories required:
▪ Demineralized water 30 °C ± 2 °C (86 °F ± 3.6 °F)
▪ Nozzle needle
▪ Brush, e.g. medium-hard toothbrush
▪ Disposable syringe
Note
Before cleaning, take the hose off the spray clip and the taper sleeve of the hand-
piece.
▶ Take the spray clip with spray tube off the handpiece to improve the cleaning.