Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF
1.011.6732
5 Operation | 5.1 Attaching the medical device
5 Operation
5.1 Attaching the medical device
CAUTION
Detachment of the medical device during treatment.
Injury or property damage
A medical device that is not properly locked in place can become disconnected from
the motor coupling and fall off.
▶ Carefully pull on the medical device before each treatment to make sure that it is
securely locked onto the motor coupling.
NOTICE
Connection to the drive motor.
Handpiece jams.
▶ Operate the handpiece only with the chuck being closed.
NOTICE
Removing and attaching the handpiece while the drive motor is rotating.
Damage to the driver.
▶ Never attach or remove the handpiece while the drive motor is rotating.
▶ Lightly spray O-rings on motor coupling with KaVo Spray.
▶ Attach the handpiece to the motor coupling until it locks into place.
▶ Before each treatment, pull on the medical device to make sure that the hand-
piece is securely seated on the motor coupling.
5.2 Removing the medical device
▶ Unlock the handpiece from the motor coupling by twisting it slightly and then
pulling it along its axis.
5.3 Inserting a straight or contra-angle bur
Note
Only use handpiece burs or contra-angle handpiece burs that are in conformance
with ISO 1797-1, type 1 and type 2, are made of steel or hard metal and meet the
following criteria:
Shaft diameter: Ø 2.334 to Ø 2.350 mm
Contra-angle burs with bur stop:
- Shaft clamping length: min. 12 mm
- Overall length: max. 22 mm
Handpiece burs without bur stop:
- Shaft clamping length: min. 30 mm
- Overall length: max. 44.5 mm