Automatic cleaning

CAUTION

Ultrasonic cleaning devices

Malfunctions from cleaning in ultrasonic cleaning devices

▶ Do not place the medical device in an ultrasonic cleaning device!

▶ Clean the ZEG tips in a thermodesinfector by Miele, for example.

▶ Use according to the manufacturer's instructions for use.

▶ After automatically cleaning the outside of the ZEG tips, immediately remove

them and sterilise them to prevent the tips from corroding.

5.2

   

Disinfection

CAUTION

Damage from liquid inside the device

If cleansers and disinfectants are improperly handled, disinfectants can enter the

device and cause malfunctions or destruction.

▶ Make sure that no disinfectants into the device

▶ Do not place the medical device in disinfectant baths.

5.2.1

   

Disinfecting the instrument

Manual disinfection

KaVo recommends the following products based on material compatibility. The mi‐

crobiological efficacy must be ensured by the disinfectant manufacturer.

Permissible disinfectants:

▪ AF liquid

▪ S&M Microcide

▪ Dürr FD 322

▶ Use disinfectants according to the manufacturer's instructions for use.

Automatic disinfecting

Not applicable.

5.2.2

   

Disinfecting the grip sleeve

Manual disinfection

KaVo recommends the following products based on material compatibility. The mi‐

crobiological efficacy must be ensured by the disinfectant manufacturer.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Instructions for use PIEZOlux (0.571.5211 - 1.001.8522)
5 Preparation methods DIN EN ISO 17664 | 5.2 Disinfection

10/15

Содержание PIEZOlux 0.571.5211

Страница 1: ...Instructions for use PIEZOlux 0 571 5211 1 001 8522 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...oad test for early recognition of tip breakage 6 4 Operation 7 4 1 Using the PIEZOlux 7 5 Preparation methods DIN EN ISO 17664 8 5 1 Cleaning 8 5 1 1 Cleaning the instrument 8 5 1 2 Cleaning the grip...

Страница 4: ...ons EMC Electro magnetic compatibility 1 1 2 Symbols Refer to chapter Safety Warning Important information for users and technicians KaVo products with this symbol can be sterilised in autoclaves up t...

Страница 5: ...m 1 3 Transportation and storage CAUTION It is hazardous to start up the medical device after it has been stored refrigerated This can cause the medical device to malfunction Prior to start up very co...

Страница 6: ...ysical harm WARNING WARNING indicates a hazardous situation that can cause death or serious injury DANGER DANGER indicates the maximum hazard level indicates a directly hazardous situation that can ca...

Страница 7: ...her brands The PIEZOlux scaler tips are intended for effective and gentle calculus removal The PIEZOlux Paro tips are used in periodontology gentle cleaning PIEZOlux cem tips are used for the insertio...

Страница 8: ...ps Insert the tip into the torque wrench and screw it into the handpiece by turning it clockwise The tip is tightly seated when the torque wrench clicks 3 2 Load test for early recognition of tip brea...

Страница 9: ...afety precaution 10 N corresponds to weight of 1 kg See also 3 2 Load test for early recognition of tip breakage Page 6 Select the spray on the foot control or the electronics Note The water is suppli...

Страница 10: ...unction properly when disinfectants are used that are listed by KaVo since they were tested for proper use on KaVo prod ucts Infection protection For treatments that are a critical treatment according...

Страница 11: ...ng CAUTION Ultrasonic cleaning devices Malfunctions from cleaning in ultrasonic cleaning devices Do not place the medical device in an ultrasonic cleaning device Clean the grip sleeve in a thermodesin...

Страница 12: ...destruction Make sure that no disinfectants into the device Do not place the medical device in disinfectant baths 5 2 1 Disinfecting the instrument Manual disinfection KaVo recommends the following p...

Страница 13: ...ased on material compatibility The mi crobiological efficacy must be ensured by the disinfectant manufacturer Permissible disinfectants AF liquid S M Microcide D rr FD 322 Use disinfectants according...

Страница 14: ...ve 140 C 284 F Do not exceed even when drying Immediately remove the parts from the steriliser after the sterilisation cycle 5 3 2 Sterilising the ZEG tips CAUTION Contact corrosion Contact corrosion...

Страница 15: ...ition of the contacts is correct Connect the hose coupling to the handpiece and screw it to the hose casing 5 4 2 Care of ZEG tips If needed open the water passage of the tips using a nozzle needle 5...

Страница 16: ...ength using the measuring card Mat no 1 001 6958 If the end of the calculus remover tips are in the red range they need to be discarded Dispose of damaged instrument inserts Dispose of worn instrument...

Страница 17: ...2 PIEZOlux Perio tip No 63 Mat no 1 000 2222 PIEZOlux Perio tip No 64 Mat no 1 000 2223 PIEZOlux Perio tip No 65 Mat no 1 000 2224 PIEZOlux cem tip No 47 Mat no 0 571 6961 Torque wrench Mat no 1 000 4...

Страница 18: ......

Страница 19: ......

Страница 20: ...1 005 5901 kb 20100125 02 en...

Отзывы: