5
Preparation methods DIN EN ISO 17664
CAUTION
Damage from improper cleaning and disinfection
Using of improper cleaners and disinfectants can restrict function or damage the
unit.
▶ Only clean the outer surface.
▶ Only use a soft cloth and tap water.
▶ Do not use solvents or aggressive chemicals.
CAUTION
Damage from liquid inside the device.
If the cleansers and disinfectants are improperly handled, liquid in the interior of
the unit can cause malfunctions or destruction.
▶ Make sure that no liquid cleanser or disinfectant enters the device.
▶ Do not place the medical device in disinfectant baths.
Note
Improper service and care can cause premature wear and malfunctioning.
KaVo only guarantees that its products will function properly when disinfectants are
used that are listed by KaVo since they were tested for proper use on KaVo prod‐
ucts.
Infection protection
For treatments that are a "critical treatment according to RKI," KaVo recommends
using an infection protection system to prevent contamination of the tip thread.
▶ Use finger stalls that are approved for medical use.
▶ Place the fingerstall over the handpiece after the tip has already been attached.
The handpiece is ready to use.
KaVo recommends the following product, for example:
▪ Disposable fingerstalls, powder-free, size: small, order number: 01271
KaVo recommends setting up the instruments as soon as they are used.
5.1
Cleaning
5.1.1
Cleaning the instrument
Manual cleaning
Required accessories:
▪ Tap water 30°C ± 5°C (86°F ± 10°F)
▪ Brush such as a medium hard toothbrush
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522)
5 Preparation methods DIN EN ISO 17664 | 5.1 Cleaning
8/15
Содержание PIEZOlux 0.571.5211
Страница 1: ...Instructions for use PIEZOlux 0 571 5211 1 001 8522 Always be on the safe side...
Страница 18: ......
Страница 19: ......
Страница 20: ...1 005 5901 kb 20100125 02 en...