background image

5.3.1

   

Sterilising the grip sleeve

Requirement

Remove the grip sleeve from the handpiece.

▶ Seal the grip sleeve in sterile goods packaging.

▶ Sterilise the grip sleeve an autoclave.

Depending on the device, select a suitable method from the following sterilisation

methods.

Autoclave with a triple pre-vacuum for least four minutes at 134°C ± 1°C (273°F ±

1.8°F)

Autoclave using the gravitational method: sterilise for at least 10 min. at 134°C ±

1°C (273°F ± 1.8°F)

Autoclave using the gravitational for at least 60 min. at 121°C ± 1°C (250°F ± 1.8°F)

Follow the manufacturer's instructions for use.

CAUTION

Overheating.

Damage to the on the grip sleeve.

▶ 140°C (284 °F) Do not exceed even when drying.

▶ Immediately remove the parts from the steriliser after the sterilisation cycle.

5.3.2

   

Sterilising the ZEG tips

CAUTION

Contact corrosion

Contact corrosion can damage the tips or cause them to break.

▶ Do not place tips in the drill bit bath or disinfection bath.

▶ Screw off the tips from the instrument using a torque wrench.

▶ Pack the tips by itself in a STERIclave bag.

Note
The STERIclave bag must be large enough so that the bag is not stretched.

▶ Sterilise calculus remover tips in autoclaves.

Depending on the device, select a suitable method from the following sterilisation

methods.

Autoclave with a triple pre-vacuum for least four minutes at 134°C ± 1°C (273°F ±

1.8°F)

 

 

 

 

 

 

 

 

 

 

 

Instructions for use PIEZOlux (0.571.5211 - 1.001.8522)
5 Preparation methods DIN EN ISO 17664 | 5.3 Sterilisation

12/15

Содержание PIEZOlux 0.571.5211

Страница 1: ...Instructions for use PIEZOlux 0 571 5211 1 001 8522 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...oad test for early recognition of tip breakage 6 4 Operation 7 4 1 Using the PIEZOlux 7 5 Preparation methods DIN EN ISO 17664 8 5 1 Cleaning 8 5 1 1 Cleaning the instrument 8 5 1 2 Cleaning the grip...

Страница 4: ...ons EMC Electro magnetic compatibility 1 1 2 Symbols Refer to chapter Safety Warning Important information for users and technicians KaVo products with this symbol can be sterilised in autoclaves up t...

Страница 5: ...m 1 3 Transportation and storage CAUTION It is hazardous to start up the medical device after it has been stored refrigerated This can cause the medical device to malfunction Prior to start up very co...

Страница 6: ...ysical harm WARNING WARNING indicates a hazardous situation that can cause death or serious injury DANGER DANGER indicates the maximum hazard level indicates a directly hazardous situation that can ca...

Страница 7: ...her brands The PIEZOlux scaler tips are intended for effective and gentle calculus removal The PIEZOlux Paro tips are used in periodontology gentle cleaning PIEZOlux cem tips are used for the insertio...

Страница 8: ...ps Insert the tip into the torque wrench and screw it into the handpiece by turning it clockwise The tip is tightly seated when the torque wrench clicks 3 2 Load test for early recognition of tip brea...

Страница 9: ...afety precaution 10 N corresponds to weight of 1 kg See also 3 2 Load test for early recognition of tip breakage Page 6 Select the spray on the foot control or the electronics Note The water is suppli...

Страница 10: ...unction properly when disinfectants are used that are listed by KaVo since they were tested for proper use on KaVo prod ucts Infection protection For treatments that are a critical treatment according...

Страница 11: ...ng CAUTION Ultrasonic cleaning devices Malfunctions from cleaning in ultrasonic cleaning devices Do not place the medical device in an ultrasonic cleaning device Clean the grip sleeve in a thermodesin...

Страница 12: ...destruction Make sure that no disinfectants into the device Do not place the medical device in disinfectant baths 5 2 1 Disinfecting the instrument Manual disinfection KaVo recommends the following p...

Страница 13: ...ased on material compatibility The mi crobiological efficacy must be ensured by the disinfectant manufacturer Permissible disinfectants AF liquid S M Microcide D rr FD 322 Use disinfectants according...

Страница 14: ...ve 140 C 284 F Do not exceed even when drying Immediately remove the parts from the steriliser after the sterilisation cycle 5 3 2 Sterilising the ZEG tips CAUTION Contact corrosion Contact corrosion...

Страница 15: ...ition of the contacts is correct Connect the hose coupling to the handpiece and screw it to the hose casing 5 4 2 Care of ZEG tips If needed open the water passage of the tips using a nozzle needle 5...

Страница 16: ...ength using the measuring card Mat no 1 001 6958 If the end of the calculus remover tips are in the red range they need to be discarded Dispose of damaged instrument inserts Dispose of worn instrument...

Страница 17: ...2 PIEZOlux Perio tip No 63 Mat no 1 000 2222 PIEZOlux Perio tip No 64 Mat no 1 000 2223 PIEZOlux Perio tip No 65 Mat no 1 000 2224 PIEZOlux cem tip No 47 Mat no 0 571 6961 Torque wrench Mat no 1 000 4...