background image

4

   

Operation

4.1

   

Using the PIEZOlux

See also: Instructions for use for the KaVo treatment unit

The KaVo ZEG PIEZOlux can gently remove calculus and plaque as well as excess

cement from natural teeth without pain at a vibrational frequency of 30 kHz.

CAUTION

Instrument inserts can be damaged from long-term use, or when dropped or bent.

They cannot be guaranteed to function properly.

Potential injury from insert breakage.

▶ Check the instrument inserts before each use.

▶ In the case of frequent use, replace the instrument inserts every 9 - 12 months.

▶ Before each use, apply approx. 10 N of pressure to instrument inserts as an

additional safety precaution. (10 N corresponds to weight of 1 kg).

See also: 3.2 Load test for early recognition of tip breakage, Page 6

▶ Select the spray on the foot control or the electronics.

Note
The water is supplied through the working tip and can be adjusted corresponding

to the tip on the water control.

The water flow should be adjusted so that the instrument tip sprays the water with

its vibration.

Water cock

▶ Control the intensity using a foot control or the electronics.

▶ Remove calculus, plaque or excess cement with light, painting movements fol‐

lowing the tooth surface.

Make sure that the movement of the tip is in line with the handpiece axis. (The

optimum removal occurs at a vibration width of 0.05 mm to 0.07 mm with a fre‐

quency of 30 kHz.)

 

 

 

 

 

 

 

 

Instructions for use PIEZOlux (0.571.5211 - 1.001.8522)
4 Operation | 4.1 Using the PIEZOlux

7/15

Содержание PIEZOlux 0.571.5211

Страница 1: ...Instructions for use PIEZOlux 0 571 5211 1 001 8522 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...oad test for early recognition of tip breakage 6 4 Operation 7 4 1 Using the PIEZOlux 7 5 Preparation methods DIN EN ISO 17664 8 5 1 Cleaning 8 5 1 1 Cleaning the instrument 8 5 1 2 Cleaning the grip...

Страница 4: ...ons EMC Electro magnetic compatibility 1 1 2 Symbols Refer to chapter Safety Warning Important information for users and technicians KaVo products with this symbol can be sterilised in autoclaves up t...

Страница 5: ...m 1 3 Transportation and storage CAUTION It is hazardous to start up the medical device after it has been stored refrigerated This can cause the medical device to malfunction Prior to start up very co...

Страница 6: ...ysical harm WARNING WARNING indicates a hazardous situation that can cause death or serious injury DANGER DANGER indicates the maximum hazard level indicates a directly hazardous situation that can ca...

Страница 7: ...her brands The PIEZOlux scaler tips are intended for effective and gentle calculus removal The PIEZOlux Paro tips are used in periodontology gentle cleaning PIEZOlux cem tips are used for the insertio...

Страница 8: ...ps Insert the tip into the torque wrench and screw it into the handpiece by turning it clockwise The tip is tightly seated when the torque wrench clicks 3 2 Load test for early recognition of tip brea...

Страница 9: ...afety precaution 10 N corresponds to weight of 1 kg See also 3 2 Load test for early recognition of tip breakage Page 6 Select the spray on the foot control or the electronics Note The water is suppli...

Страница 10: ...unction properly when disinfectants are used that are listed by KaVo since they were tested for proper use on KaVo prod ucts Infection protection For treatments that are a critical treatment according...

Страница 11: ...ng CAUTION Ultrasonic cleaning devices Malfunctions from cleaning in ultrasonic cleaning devices Do not place the medical device in an ultrasonic cleaning device Clean the grip sleeve in a thermodesin...

Страница 12: ...destruction Make sure that no disinfectants into the device Do not place the medical device in disinfectant baths 5 2 1 Disinfecting the instrument Manual disinfection KaVo recommends the following p...

Страница 13: ...ased on material compatibility The mi crobiological efficacy must be ensured by the disinfectant manufacturer Permissible disinfectants AF liquid S M Microcide D rr FD 322 Use disinfectants according...

Страница 14: ...ve 140 C 284 F Do not exceed even when drying Immediately remove the parts from the steriliser after the sterilisation cycle 5 3 2 Sterilising the ZEG tips CAUTION Contact corrosion Contact corrosion...

Страница 15: ...ition of the contacts is correct Connect the hose coupling to the handpiece and screw it to the hose casing 5 4 2 Care of ZEG tips If needed open the water passage of the tips using a nozzle needle 5...

Страница 16: ...ength using the measuring card Mat no 1 001 6958 If the end of the calculus remover tips are in the red range they need to be discarded Dispose of damaged instrument inserts Dispose of worn instrument...

Страница 17: ...2 PIEZOlux Perio tip No 63 Mat no 1 000 2222 PIEZOlux Perio tip No 64 Mat no 1 000 2223 PIEZOlux Perio tip No 65 Mat no 1 000 2224 PIEZOlux cem tip No 47 Mat no 0 571 6961 Torque wrench Mat no 1 000 4...

Страница 18: ......

Страница 19: ......

Страница 20: ...1 005 5901 kb 20100125 02 en...

Отзывы: