Instructions for use KaVoLUX 540 LED U / T
8 Safety check - Test instructions | 8.1 Mounting the device
92 / 110
8 Safety check - Test instructions
8.1 Mounting the device
Note
The test instructions for the KaVoLUX 540 LED U / T to be mounted to
devices are described in the instructions for use of the respective treatment
unit
The measurement of the protective earth, defined in the instructions for use
for the relevant treatment unit, must be extended to include the measuring
points "Scan the light head of the treatment light KaVoLUX 540 LED with the
probe".
See also:
2
8.2.2.2.4.2 Scan the light head of the treatment light KaVoLUX 540 LED
with the probe, Page 101
8.2 Ceiling mounting
Note
This chapter describes the safety check performed on the
KaVoLUX 540 LED T for the case of ceiling mounting.
8.2.1 Introduction
General instructions
Note
The safety check may only be carried out by one or more electricians (as
defined in IEC 61140) who have been appropriately trained for the device to
be inspected.
Note
The contents and specified tests described in this document are based on the
international standard, IEC 62353. This standard applies to the testing and
inspections of medical electrical devices or medical electrical systems comply-
ing with IEC 60601-1 (DIN EN 60601-1).
Note
In order to evaluate the safety of medical devices, systems or components of
medical devices or systems, the safety check must be carried out at the fol-
lowing times:
▶
Prior to startup
▶
during servicing
▶
during inspections and maintenance
▶
following repairs
▶
on the occasion of recurrent tests
Содержание LUX 540 LED U
Страница 1: ...Instructions for use KaVoLUX 540 LED U T ...
Страница 109: ......
Страница 110: ...1 008 8651 Fk 20190603 12 en ...