4 Assembly | 4.4 Assembly of the KaVoLUX 540 LED T
4.4 Assembly of the KaVoLUX 540 LED T
4.4.1 ESTETICA E30
WARNING
Electrical power.
Death or injury from electric shock.
▶
Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
▶
After conversion, check the electrical safety according to DIN EN 62353
[IEC 62353].
Install the mounting pole
The patient chair must be moved downwards to install the operating light.
▶
Press button "left" and "right" on the foot control simultaneously and press
down the cross switch to move the patient chair downwards.
1
6
3
4
1
5
2
▶
If the fastening screws
⑥
penetrate the mount, lift the cover
⑤
and turn
the screws back.
▶
Pick up the insertion tool
①
at the receiver tube and push through the lamp
mounting pole
③
.
▶
Insert the lamp mounting pole
③
into the receiver tube as far as it will go.
▶
Align the lamp mounting pole so that the mounting hole
②
is pointing in the
12 o'clock direction (to the head end).
▶
Fasten the lamp mounting pole with three fastening screws
⑥
.
▶
Tighten the PE screw
④
.