![KaVo LUX 540 LED U Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/kavo/lux-540-led-u/lux-540-led-u_instructions-for-use-manual_3747369052.webp)
Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.4 Assembly of the KaVoLUX 540 LED T
52 / 110
9
10
Operating light automatic, On/OFF function via the AP positions of the treat-
ment centre:
The operating light is connected via the DCA relay.
Note
ERGOcom light may not be connected to the DCA relay (X28) if this connec-
tion is possible.
ERGOcom light will be damaged otherwise.
▶
Connect cable
①
to terminal strip 24 V AC .
▶
Connect cable
②
to DCA relay, slot 12
④
.
▶
Route a wire jumper from the DCA relay, slot 13
⑤
to terminal strip 0 V AC
⑥
.