8
Malfunctions
8.1
Error messages
The user guidance system controlled by a microprocessor helps you operate the
KEY Lasers 3+ / III upgraded with a touchscreen. In addition, there are numerous
automatic self tests incorporated in the unit that control the device.
For example, lamp age and misadjustment are automatically compensated by
means of a microprocessor controlled energy regulation with self adjustment.
If the microprocessor determines malfunctions, they are located and shown on the
touchscreen.
A distinction is drawn between warnings and malfunctions.
Warnings are displayed on the touchscreen with the word WARNING, and mal‐
functions are displayed with the word MALFUNCTION.
8.2
Warning
If the spray water supply is becoming empty during operation, the message on the
touchscreen only appears after leaving Ready mode.
Press the continuing key to continue working Ready mode.
See also: 4.2 Fill the spray water, Page 30
Note
The spray reservoir should be refilled as soon as possible since the spray water
level will only be checked the next time the device is turned on.
8.3
Malfunctions
The unit automatically turns off 30 seconds after a malfunction is displayed on the
touchscreen.
▶ Turn the unit on again.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
8 Malfunctions | 8.1 Error messages
70/76