4.9.3
S-PULSE (short pulses)
By using the S-PULSE function, you can switch to short pulses when working on
hard tissue (programs 15 - 18) and thereby increase the individual pulse efficiency.
S-PULSE mode is only useful for processing dental hard tissue and is not required
for processing softer materials (such as calculus and soft tissue).
S-Pulse is not activated.
S-Pulse is activated.
Note
S-PULSE can only be selected for programs 15 to 18.
When creating your own programs, S-PULSE can only be saved in combination
with the laser contra-angle handpiece 2060 and the laser contra-angle handpiece
K 2063.
4.9.4
Detection mode
The detection mode is for detecting tissue.
If the handpiece is guided over a tooth, the following display appears:
▪ The momentary measured fluorescence appears below Mom.
▪ The maximum value measured since the last reset appears below PEAK.
The peak value can be reset to the current momentary value in detection mode
by pressing the laser pulse trigger.
▪ The last entered threshold appears below the threshold key.
See also: 4.10 Setting the threshold, Page 44
4.9.5
Feedback mode
Feedback mode combines the detection function in detection mode with the thera‐
peutic function of the laser beam.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
4 Operation | 4.9 Modes
43/76