Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
Note
Additional measuring points AP X must be taken into consideration in the
presence of accessories: e.g. accessories such as PIEZO ultrasonic scaler,
etc.
See also:
2
8 Annex - Additional measuring points, Page 111
Connect accessible conductive parts [ACP] with PE
ACP = accessible conductive parts
ACP
ACP
Note
Additional measuring points ACP X must be taken into consideration in the
presence of accessories.
See also:
2
8 Annex - Additional measuring points, Page 111
ACPs on the treatment centre
No ACPs need to be connected to the protective conductor (PE) during the
measurement on the treatment centre Primus 1058 Life, as all relevant parts
are connected to the PE at the factory and are included in the test.
ACPs on treatment lamps
No ACPs need to be connected to the operating lights during the measurement
with protective conductor (PE) because all relevant parts have already been
connected to the protective conductor (PE) at the factory and are included in
the test.
Measure protective conductor resistance
Limit
< 0,3 Ω (maximum value!)
Note
The integrity of the power supply cable, in particular the protective earth wire
of the power cable must be ensured. As this is a fixed installation, the evalu-
ation can be conducted by means of a visual inspection. If damage is determ-
ined, the further procedure to be taken is specified in the general instruc-
tions.
99 / 126