Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
94 / 126
7.2.2 Visual inspection (inspection by examination)
Check the following items in advance:
▪ Has the equipment of the ME device or the ME system been changed since
the last inspection?
▪ Has the change been documented and approved (test protocol of safety
check STK)?
▪ Are there any indications of insufficient safety?
Check the ratings of fuses that are accessible from
outside
Note
A check of whether or not the main fuse on the mains socket corresponds to
the rated values is no longer included in DIN EN 62353:2015 [IEC
62353:2015].
Visual inspection and appraisal of the medical device
and accessories
The following list is for exemplary purposes and makes no claim of being com-
plete.
Check the following items:
▪ Stability of the device
▪ No damage to the cladding or casing (cracks, breakage)
▪ Functioning of the carrier systems on dentist and assistant side, treatment
lamp, and display (brakes, height adjustment, etc.)
▪ Condition of the handpiece and suction hoses
▪ Condition of all installed application parts
▪ Condition of the control panels
▪ Condition of the threads for the fitting of tips to the ultrasound scaler hand-
piece
▪ Condition of the operating light
▪ Absence of leaks on the body of the device
▪ Condition of the power connection provided by the treatment centre