7 Safety check - Test instructions | 7.2 Instructions for the safety check
100 / 126
Note
In this measurement the resistance of the protective earth connection of the
supply network can be taken into consideration.
Note
If applicable: all removable supply connection lines, which are kept handy for
possible use, should be taken into consideration and the respective PE meas-
ured.
Protective earth measurement
The protective conductor resistance must be measured at the following parts of
the device:
▪ Treatment centre
▪ Treatment lamp
▪ Accessories
Note
Additional measuring points SL X need to be taken into consideration in the
presence of accessories: e.g. additional devices, such as third-party con-
nector, USB port of the intraoral camera, etc.
See also:
2
8 Annex - Additional measuring points, Page 111