Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
104 / 126
Scan the treatment lamp with the test tip
Operating light KaVoLUX 540 LED U
①
Fastening screw of the handle sup-
port when the gripping sleeve has
been removed
Operating light EDI/MAIA
No measuring points need to be scanned on the operating lights EDI and MAIA.
Touch monitor with test tip:
▶
Scan the measuring point
①
with the test tip.
or
▶
Scan the measuring point
②
after removing the display cover.
Measure protective conductor resistance of accessories
See also:
2
8 Annex - Additional measuring points, Page 111