Instructions for use HealOzone 2130 C
6 Preparation methods DIN EN ISO 17664
|
6.1 Cleaning
6.1.2 Clean handpiece
▪ Disinfect or sterilise handpiece before first use.
▪ A new silicone cup must be used for every patient.
▪ Silicone cups are disposable and cannot be disinfected; they must therefore be
used for treating one patient only.
▪ The handpiece must be cleaned and disinfected before each use and between
every application.
Manual cleaning of the exterior
▶ Clean the exterior of the product by brushing it off under flowing water until there
is no more visible soiling.
Interior cleaning by hand
▶ Connect up the handpiece and instrument tubing.
▶ Place handpiece in a drip pan or basin ①.
▶ Fit adapter ③ onto quick coupling ② and rinse for 15 seconds using a water gun
with cold tap water.
or
▶ Attach a standard 50ml infusion syringe and perform ten 50ml rinses.
2
3
1
Automated external and internal cleaning
Using a Miele thermodisinfector, for instance. Use in accordance with
manufacturer's user instructions. Dry immediately afterwards to ensure product re‐
mains fully functional.
40/57