Instructions for use HealOzone 2130 C
4 Commissioning
|
4.2 Connections
4.2 Connections
4.2.1 Connect up the handpiece and instrument tubing
The handpiece and instrument tubing are equipped with a quick coupling.
Note
The coupling at the end of the tubing is identical and can be connected in both
directions.
▶ Make sure there is an O-ring on the coupling and that it is undamaged.
▶ Grease the coupling O-ring with silicone grease Mat. no. 1.000.6403.
▶ Fit handpiece onto coupling, making sure that it snaps in place.
▶ Give the handpiece a gentle tug to check it is securely in place.