Instructions for use DIAGNOcam Vision Full HD
9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 9.2 Immunity to elec-
tromagnetic interference
9.2 Immunity to electromagnetic interference
The DIAGNOcam Vision Full HD is designed for use in an environment of the
kind specified below. The customer or the user of the DIAGNOcam Vision Full
HD should ensure that the device is used in an environment of the specified
type.
Interference im-
munity tests
EN 60601 test level
Compliance level
Electromagnetic en-
vironment -
Guidelines
Electrostatic discharge
(ESD) according to EN
61000-4-2
± 8 kV contact discharge
± 15 kV atmospheric
discharge
± 8 kV contact discharge
± 2/4/8/15 kV atmo-
spheric discharge
Floors should be made of
wood or concrete or be
fitted with ceramic tiles.
If the floor is fitted with
synthetic material, the
relative humidity must
be at least 30 %.
Fast transient electrical
interference / bursts ac-
cording to EN 61000-4-4
± 1 kV for input and out-
put lines
± 1 kV for input and out-
put lines
The quality of the supply
voltage should corres-
pond to that of a typical
business or
hospital environment.
Magnetic field at a sup-
ply frequency (50/60 Hz)
according to EN
61000-4-8
30 A/m
30 A/m
Magnetic fields at the
mains frequency should
correspond to typical
values in a business and
hospital environment.
Note: U
T
is the alternating mains voltage before the test level is used.